Assessment of tube current modulation in pelvic CT - BIR Publications

14 downloads 0 Views 177KB Size Report
to noise ratio was calculated using Polytetrafluoroethylene (PTFE) and distilled water inserts; PTFE was used to represent bony structure and distilled water was ...
The British Journal of Radiology, 79 (2006), 62–70 DOI: 10.1259/bjr/50019934

E

2006 The British Institute of Radiology

Assessment of tube current modulation in pelvic CT G R IBALL, MSc, DipIPEM, D S BRETTLE, PhD and A C MOORE, MSc, DipIPEM Department of Medical Physics & Engineering, Leeds Teaching Hospitals NHS Trust, Leeds General Infirmary, Leeds LS1 3EX, UK

Abstract. An anatomically shaped polymethylmethacrylate (PMMA) phantom was used to assess the effect of the Siemens CARE Dose mA modulation system on pelvic CT scans. The effect of the system on absorbed dose to air, image percentage noise and the signal to noise ratio of clinically relevant details was assessed. The signal to noise ratio was calculated using Polytetrafluoroethylene (PTFE) and distilled water inserts; PTFE was used to represent bony structure and distilled water was used to represent soft tissue abscess. Pelvis protocols identified from local hospitals and the UK CT Dose Survey (2002), were assessed and compared with those provided by Siemens Medical (UK). These protocols were tested on a Siemens Sensation 4 CT scanner, both with and without CARE Dose. Results were obtained which showed that dose savings were possible with no significant increase in image noise. Dose reductions were 8% in the lateral positions in the phantom and 42% in the centre, top and bottom. The calculated ‘‘CTDIvol’’ was 32% lower with CARE Dose than without CARE Dose. This is slightly greater than the 25% change in the effective mAs values that was found. This implies that the reduction in the effective mAs values is a reasonable predictor of the total reduction in absorbed dose to air, whilst slightly underestimating the actual change. The results also showed a non-significant trend towards decreased signal to noise ratios for clinically relevant CT numbers when CARE Dose was activated. This suggests that tube current modulation may detrimentally affect signal detection due to changes in image noise.

CT examinations account for a large proportion of the collective dose from medical X-ray examinations in the UK. In 2000 this was reported as being 40% [1] but may now be even higher due to an increase in the range and volume of routine examinations and the uptake of CT fluoroscopy and cardiac CT scanning. There is a requirement for all X-ray examinations to be optimized such that the patient dose is ‘‘As Low As Reasonably Practicable’’ (ALARP) [2]. However, it is often difficult to implement procedures which significantly lower the radiation dose without decreasing the image quality to a non-diagnostic level. One recent technological advance from CT manufacturers in terms of dose reduction has been to introduce tube current modulation systems for CT scanning. The approach taken by Siemens Medical Systems (Erlangen, Germany) is a system called CARE Dose which claims to reduce patient doses whilst having no significant adverse effects on the image quality. This system has been described extensively in the literature [3–5]. The human body varies in composition both along its length and in the transverse plane at any given point along the body. This produces variations in X-ray attenuation due to both the external dimensions of the body and its internal composition. In CT scanning, as the X-ray tube and detectors rotate around the body, the attenuation can change by two orders of magnitude [4]. These differences in attenuation are most significant in the regions of the shoulder and pelvis, where large thicknesses of bone are found in the lateral projections, but a much smaller thickness of bone is present in the anterior–posterior projections. It is these examinations which provide the greatest challenges, in terms of the dose–image quality balance. As a result, using a constant tube current (mA) for each scan angle within a given rotation may result in Received 18 October 2005 and in final form 24 May 2005, accepted 31 May 2005.

62

either photon starvation artefacts on the high attenuation projections or overdosing in the lower attenuation projections. In the CARE Dose system, during each rotation of the tube and detector assembly around the patient, a small number of the central detector channels provide attenuation information, which is dependent upon the patient cross section and scan angle, to the X-ray generating system [3]. The information provided by these detector channels is used to determine to what extent the mA can be modulated, with respect to an initial tube current setting, without adversely affecting the image quality. As a result the tube current is modulated dynamically with a delay of one rotation relative to the attenuation measurement. The first patient based assessment by Greess et al [6] showed that, when CARE Dose is used, a dose reduction of approximately 25% (in terms of total mAs reduction) is possible in pelvic scanning ‘‘with no significant decrease’’ in subjective assessments of image quality. Similar percentage dose reductions have been demonstrated in other clinical work [7] and these showed good agreement with phantom based data [2, 5]. Most of the published work has used image noise and/or subjective image assessment to quantify image quality. A small number of papers [8, 9] have used standard deviations from regions of interest (ROIs) to yield a more objective assessment of image noise. Claims that the image quality was not affected by the CARE Dose system were queried by local users. Having used CARE Dose for a period of time, they perceived that the quality of the images for pelvis scans was subjectively worse when CARE Dose was used and this raised concerns that it may have a detrimental effect on the accuracy of diagnosis. This is despite the manufacturer’s recommendation that CARE Dose is used for all clinical situations other than for extremely large patients. This discrepancy between the reported claims and The British Journal of Radiology, January 2006

Assessment of tube current modulation

local experience prompted this investigation into the relationships between patient dose, image percentage noise and the signal to noise ratio (SNR) as an indicator of diagnostic detectability. The objective was to clarify whether the CARE Dose system can yield significant dose reduction for no loss of image quality in pelvic scanning.

Materials and methods A series of measurements were made using an anatomically shaped polymethylmethacrylate (PMMA) phantom which has been described in the literature [3] as a ‘‘hip’’ phantom. A schematic diagram of the phantom is shown in Figure 1. The thickness of the phantom is 14.5 cm in the z-axis. The hip phantom does not contain any bony structure and therefore the similarity of the phantom to the pelvic region is geometric only. As such it may be expected that the magnitude of the tube current modulation in clinical practice may be different from that found for this phantom.

Protocol selection Routine protocols for soft tissue assessment of the pelvis on Siemens 4 slice CT scanners (Somatom Volume Zoom and Somatom Sensation 4) were obtained from three local hospitals, the UK CT Dose survey 2002 and from Siemens Medical (UK) (Bracknell, UK). These protocols fell into two main groups, those that used a pitch of 1.00 and those that used a pitch of 1.25. There was some variation in the mA/mAs setting that was used, but all of the protocols used 120 kV and a rotation time of 0.5 s. In light of these findings all measurements were performed at the standard exposure factors given in Table 1. Reconstruction kernel B40s was used. Effective mAs is defined as the tube mAs per rotation divided by the helical pitch, where the pitch is the ratio of

Figure 1. A schematic diagram of the hip phantom. Solid cylinders within the phantom (1) represent the five ion chamber positions, dashed circles within the phantom (2) represent the five polymethylmethacrylate (PMMA) CT number and noise measurement positions; the dashed circle outside the phantom (3) represents the air CT number measurement position.

the table feed per rotation and the total X-ray beam width [10]. The effective mAs value of 165 was chosen as this was representative of most of the protocols that were obtained. Pitch settings of 1.00, 1.13 and 1.25 were used, both with and without CARE Dose. All the measurements were performed using the same PMMA hip phantom on the same scanner, a Siemens Sensation 4. Measurements were made which investigated how four different parameters changed with the application of CARE Dose. The four parameters that were investigated were (i) absorbed dose to air, measured in the phantom, (ii) image percentage noise, (iii) CT number for water and polytetrafluoroethylene (PTFE), and (iv) the SNR for both of these materials. These parameters were investigated for each pitch setting, both with and without CARE Dose. Water was chosen to represent low density abscess and PTFE to represent bony structure.

Consistency tests Prior to the testing, all of the test equipment was placed in the scanning room for at least 4 h in order for the temperature of the phantom and test equipment to stabilize with the room temperature. At the start of each visit the scanner was air-calibrated using the software on the scanner. A short series of consistency tests were performed immediately after the air-calibration which, on all subsequent visits, enabled us to verify that the performance of the scanner had not changed from the previous visit. On each occasion the hip phantom was positioned 15 cm from the end of the couch on top of the mattress in order to maintain consistent scattering conditions. The phantom was aligned using the laser lights on the scanner and with a spirit level. The set up is shown in Figure 2. A Scan Projection Radiograph (Topogram) of the phantom was acquired and a helical acquisition was planned from this image. The scan length for the helical acquisition was the whole length of the phantom

Figure 2. The phantom as positioned for the dose and noise measurements.

Table 1. Standard exposure factors for all scans Tube voltage (kV)

Effective mAs

Rotation time (s)

Beam collimation (mm)

Image slice thickness (mm)

SFOV (mm)

120

165

0.5

462.5

5

380

SFOV, field of view.

The British Journal of Radiology, January 2006

63

G R Iball, D S Brettle and A C Moore

(145 mm), which gave a total scan time of 8.2 s, using a pitch of 1.25. All of the consistency tests were performed with CARE Dose on. The first test was a measurement of the absorbed dose to air. A scan was performed at the standard exposure factors at a pitch of 1.25 with a calibrated 3 cm3 pencil ionization chamber (Capintec Inc., Ramsey, NJ), having an active length of 100 mm, in the central position. The chamber was connected to a Keithley 35050A Dosimeter (Keithley Instruments Inc., Cleveland, OH). The absorbed dose to air was recorded and the mean PMMA CT number and standard deviation (p) were measured adjacent to each of the five possible chamber positions (see Figure 1), on the CT slice closest to the centre of the phantom, using the region of interest (ROI) tool on the scanner. The size of the ROI that was used was kept constant throughout all of the measurements. The mean CT number of air was also measured at a standard position outside the phantom using a ROI of the same size. This scan and measurement procedure was then repeated with the ion chamber in the right lateral measurement position. On each occasion the ambient air temperature and pressure were measured, in addition to the phantom temperature, so that an air density correction could be applied to the dose measurements. The ion chamber was then removed from the phantom and a PTFE rod was inserted into the central measurement position. The scan was repeated and the mean CT number and p of the PTFE rod were recorded in addition to the measurements described above. Again this was repeated with the PTFE rod in the right lateral position.

Absorbed dose to air and noise measurements Absorbed dose to air measurements were made for each of the five chamber positions both with and without CARE Dose at each of the three pitch settings. For each scan the measured dose and total mAs were recorded. On the central slice the mean PMMA CT number and p were recorded at each measurement position and the mean air CT number was also recorded. For each measurement position the image percentage noise was calculated using Equation (1) [11]: Image Percentage Noise~

pPMMA  100 CTPMMA {CTAir

that was used in this study. However, the CTDIvol method is an accepted way of accounting for the distribution of dose within a phantom. Since, in this case, it is the comparison between the CTDIvol values for two different scanning situations, rather than the absolute value that was of most importance, the CTDIvol was used simply as an indicator of the relative change in absorbed dose to air. As such, the term ‘‘CTDIvol’’ is used for all calculations that relate to the hip phantom. The effect of the CARE Dose system on the percentage dose reduction was also evaluated over a range of initial effective mAs settings (50–200 mAs).

Signal measurements Two sets of signal to noise measurements were made, for the water and PTFE inserts. For the water measurements thin rubber sheaths were inserted into each of the five holes in the phantom and distilled water was inserted into each of the sheaths and the ends were secured with plastic clips. The sheaths were similar in diameter to the holes in the phantom which made it possible to almost completely fill the holes with water. The set up of the phantom for the water measurements is shown in Figure 3. For the PTFE measurements each individual rod was manufactured in house from a single PTFE rod (Barkston Plastics Ltd, Leeds, UK). All five rods were manufactured from the same batch of PTFE to ensure that there was no difference in composition between the individual rods. For each set of measurements the phantom was scanned five times at each pitch setting with and without CARE Dose. For each scan the mean CT number and p of the water/PTFE and PMMA were recorded at each measurement position on the central slice in addition to the mean CT number of air at the standard position. Measurements were also repeated 10 times on one scan of the PTFE rods in order to establish the repeatability of the measurements.

ð1Þ

where: p is the standard deviation and CT is the mean CT number (Hounsfield Unit) of the indicated material. The absorbed dose to air was corrected for ambient temperature and pressure and the ion chamber calibration factor was applied. The volume averaged CT dose index (CTDIvol) was then calculated for the scans with and without CARE Dose, using Equation (2) [11]. This was performed for each pitch setting:  CTDIvol ~ 1=3  CTDIcentre z2=3  CTDIperiphery =pitch ð2Þ where: CTDIcentre is the CTDI measured in the centre of the phantom and CTDIperiphery is the average of the four CTDI values which were measured in the periphery of the phantom. Pitch is the ratio of the table feed per rotation and the total X-ray beam width. CTDIvol is actually defined for a cylindrical phantom and as such it is not strictly applicable to the hip phantom 64

Figure 3. The phantom as set up for the water signal to noise ratio (SNR) measurements. The British Journal of Radiology, January 2006

Assessment of tube current modulation

The SNR for the inserts was calculated Equation (3). CTsignal {CTPMMA SNR~ pPMMA

using

ð3Þ

Where: CTsignal is the mean CT number of water or PTFE and p is the standard deviation. The modulus was used as the mean CT number for water was sometimes below zero. SNR calculations were performed for each measurement point for each pitch setting. Error propagation was performed for all of the parameters of interest and the calculated values are shown with the results. The pooled standard deviation of the SNRs was calculated for each pitch setting for the water measurements and this result was used to power the study. The powering process showed that for a result to be statistically significant at the 95% level 25 measurements were required (both with and without CARE Dose). As a result, a further set of SNR measurements were made for both water and PTFE. The phantom was set up as described earlier and 25 scans were performed both with and without CARE Dose. For each scan the mean CT number and p of the insert (PTFE/water) was measured in the central position in addition to the mean CT number and p of the PMMA adjacent to the central insert. SNRs were calculated from these measurements and errors were calculated as for the previous measurements. Statistical analysis was performed on these results (Kruskal–Wallis non parametric test) to determine whether the SNRs of water and PTFE changed significantly for the scans with CARE Dose.

Dose measurements For the scans without CARE Dose (i.e. constant mA) the absorbed doses to air were significantly higher in the top and bottom positions than in the lateral positions. For the scans with CARE Dose there was a significant decrease in the absorbed dose to air in each position. The reductions were approximately 42% in the central position, 42% in the top and bottom positions and 8% in the lateral positions. These results were as expected and are shown in Figure 4. The calculated value of ‘‘CTDIvol’’ with CARE Dose was 32% lower than the value for the scans without CARE Dose. The scanner indicated reduction in effective mAs for the scans with CARE Dose was 25% (relative to the constant tube current case). The percentage reduction in ‘‘CTDIvol’’ was independent of pitch to within 0.5% over the pitch range of 1–1.25, as shown in Figure 5. The error bars that are shown in Figure 5 represent one standard deviation about the mean. The percentage reduction in absorbed dose to air in the central position for varying initial effective mAs settings is shown in Figure 6. The reduction in absorbed dose to air is approximately 40% for mAs settings between 50 mAs and 165 mAs. However, this reduction in absorbed dose to

Results The results of the consistency tests that were performed showed that on each occasion the performance of the scanner had not changed since the first visit. For clarity all the results for the 1.25 pitch setting are shown with summary results for the other pitches.

Figure 5. Percentage reduction in ‘‘CTDIvol’’ against pitch setting.

Figure 4. Variation of absorbed dose to air with position in phantom.

The British Journal of Radiology, January 2006

65

G R Iball, D S Brettle and A C Moore

Noise measurements

Figure 6. Reduction in absorbed dose to air in the central position against initial effective mAs setting.

air rises to 50% at 200 mAs. It was not possible to obtain results for scans with mAs settings above 200 mAs as this would have exceeded the maximum tube loading at this pitch setting.

The initial set of image percentage noise values, calculated using Equation (1), across the five positions showed that, in general, there was not a large difference between the measured values with and without CARE Dose, other than for the top and bottom positions, as shown in Figure 7. These discrepancies between the noise values in the top and bottom positions were not found for the other pitch settings and are thought to be anomalous results, relative to the other pitch settings. As expected the noise values in the top and bottom were slightly different from those found in the lateral positions. For all pitches the highest noise values were found in the centre of the phantom. There was a general reduction in the image percentage noise as the pitch setting was increased (Figure 8), for both CARE Dose on and off, although this was not greater than the experimental uncertainties. For the 25 additional scans the image percentage noise was assessed in the centre of the phantom as for the initial tests. The difference between the noise values for CARE Dose on and off, which was approximately 10%, was tested for significance using the Kruskal–Wallis test. The mean and standard deviations of the noise values for the

Figure 7. Variation of image percentage noise with position in the phantom.

Figure 8. Image percentage noise variation with pitch setting.

66

The British Journal of Radiology, January 2006

Assessment of tube current modulation

25 additional scans are calculated p-values. These results show difference, at the 95% scans with and without

shown in Table 2 along with the that there was no significant level, in the noise levels for the CARE Dose.

SNR measurements The measurements for the water and PTFE inserts were used to calculate the SNR for each material (Equation (3)) and these results, for a pitch of 1.25, are shown in Figures 9 and 10. These results from the initial tests show that the SNRs, for both water and PTFE, are lowest in the centre

Table 2. Calculated mean, standard deviation and p-values for the image percentage noise tests

PTFE Water

Mean noise (SD) for CARE Dose off

Mean noise (SD) for CARE Dose on

p-value

19.4 (3.7) 17.3 (3.1)

19.0 (3.6) 19.1 (3.8)

0.727 0.099

SD, standard deviation; PTFE, polytetrafluoroethylene.

of the phantom both with and without CARE Dose. This is as expected as the noise values were highest in the centre of the phantom. The differences seen between the PTFE SNR values with and without CARE Dose, were generally within the experimental uncertainties. There appears to be a general decrease in the water SNR at each position for the scans with CARE Dose, which is an undesirable trend. However, the differences in water SNR were also within the experimental uncertainties. Figures 9 and 10 show that there are positional variations in the SNR within the phantom. As such it is not valid to average the SNR for the five different positions as this will mask the positional variations and will result in large uncertainties in the results. For a pitch of 1.25, the water SNR values measured with CARE Dose were lower than those without CARE Dose. This trend was observed for the other pitch values for the water scans but was not observed for the PTFE scans. For the 25 additional scans the signal and noise values were measured for PTFE and water in the central position in the phantom. From these results the SNR for both inserts were calculated as for the original scans. The differences between the values of CT number and SNR for the scans with CARE Dose on and off were tested for Figure 9.

Signal to noise ratios (SNR) for polytetrafluoroethylene (PTFE) for each phantom position.

Figure 10. Signal to noise ratios (SNR) for water for each phantom position.

The British Journal of Radiology, January 2006

67

G R Iball, D S Brettle and A C Moore

significance using the Kruskal–Wallis test. The mean and standard deviations of the CT numbers and SNRs and the resulting p-values are shown on Table 3. These results show that there were no significant differences, at the 95% level, in CT number or SNR between the scans with and without CARE Dose despite the SNRs generally being decreased when CARE Dose was used.

Discussion Significant reductions in absorbed dose to air were found in all five positions in the phantom for the scans with CARE Dose relative to the constant tube current situation (Figure 4). The largest reductions, up to 42%, were found in the top, bottom and central positions as these positions lie on the lowest attenuation paths through the phantom and therefore experience the largest tube current modulation and reduction in absorbed dose to air. The dose reductions in the lateral positions, around 8%, are much smaller in magnitude as the attenuation is at its highest in these positions which means the tube current will be at its maximum value. These dose reductions were smaller than those shown by Kalender et al [5] who found a 45% ‘‘average’’ dose reduction via direct dose measurement. Kalender used a scan time of 1 s (compared with a 0.5 s scan time in our work) which allowed for a larger modulation amplitude and therefore a greater dose reduction than in this study. We also found that the reduction in the effective mAs values were lower than those found by Kalender et al, at approximately 25% compared with 40%. However, Kalender’s work showed an associated increase in image noise of approximately 10%, which was not found in the first part of this study. Kalender’s work was performed with a prototype version of the CARE Dose system which may also explain some of the differences between those initial results and the results of this study. Kalender et al [5] measured an average dose reduction of 45% in the hip phantom with a 3 cm3 ionization chamber similar to that used in this study. This 45% reduction in dose, however, was a straightforward average of the five measurement points rather than a volume average (‘‘CTDIvol’’) which was calculated here. A straightforward average of our results yields a dose reduction of 30%. The reduction in ‘‘CTDIvol’’ of approximately 32% was in good agreement with the relative dose reduction found by Gies et al [4], who found dose reductions of approximately 38%, for computer simulations using the hip phantom. The large reduction in absorbed dose to air in the central position is of importance as most of the more radiosensitive organs lie centrally. These results imply that the reduction in an individual organ dose (with an associated change in the effective dose) may be larger than the reduction in the values of ‘‘CTDIvol’’ shown

here. These results have implications for calculating effective doses in CT as the current Monte Carlo data sets that are used do not reflect the distribution of dose within the patient when a tube current modulation system is used. The large dose reduction in the centre of the phantom also has significant implications for pelvic scans of pregnant patients. If CARE Dose was used for these patients the risk to the fetus may be significantly reduced relative to scans performed with a constant tube current. Tack et al [10] showed that when using CARE Dose, the percentage dose reduction was independent of the initial effective mAs setting. They used six different mAs settings between 20 mAs and 100 mAs for chest and abdomen CT scans. Our results (Figure 6) show that the percentage dose reduction is approximately constant at a value of around 40% for initial effective mAs values up to 165 mAs. Above this value the percentage dose reduction increases, to approximately 50% at 200 mAs. This occurred as the mAs setting approached the maximum tube current rating for the tube. The Manufacturers recommend that for extremely large patients, where the mAs setting may be close to the tube limit, CARE Dose is not used. No measurements were made to determine whether or not the tube output varied linearly with mAs so we cannot exclude poor output linearity with mAs as a possible cause of the results shown in Figure 6. The image percentage noise level was not significantly affected by the application of CARE Dose, as shown in Figure 7, for the initial set of noise measurements. The reduction in dose of approximately 8% in the left and right positions occurs as a result of the integration of the reduction in tube current over all scan angles as there is no reduction in the tube current setting in the lateral projections. Given that there has been a general reduction in dose across the phantom there should have been an associated increase in the image percentage noise. No such increase in image percentage noise was found. Combining these results and those for the dose measurements shows that the reductions in absorbed dose to air that were calculated are net dose savings, i.e. they come with no significant noise penalty. Previous work [3, 5–7] showed that dose reductions of 23–45% were possible in the pelvis region with no significant difference in subjective assessments of image quality. The slight decrease in the image percentage noise with pitch setting, for both CARE Dose on and off is thought to be due to the combined effect of setting a constant effective mAs value and the magnitude of the over-scan which is necessary in helical scanning. For the additional scans with the water inserts there was an increase in the noise level of approximately 10% for the scans with CARE Dose on relative to the scans with CARE Dose off. This was not found to be significant at the 95% level (p50.099). This 10% increase in noise agrees

Table 3. Calculated mean, standard deviation and p-values for CT number and SNR for water and PTFE

PTFE Water

Mean CT number (SD) for CARE Dose off

Mean CT number (SD) for CARE Dose on

Mean SNR (SD) for CARE Dose off

Mean SNR (SD) for CARE Dose on

p-value (CT number)

p-value (SNR)

962.4 (4.6) 7.0 (3.6)

961.2 (4.5) 6.5 (3.3)

44.3 (8.8) 7.6 (1.7)

44.9 (8.8) 7.0 (1.6)

0.393 0.421

0.764 0.197

SD, standard deviation, SNR, signal to noise ratio, PTFE, polytetrafluoroethylene.

68

The British Journal of Radiology, January 2006

Assessment of tube current modulation

well Kalender’s work [5]. A similar change in noise was not found for the scans of the PTFE inserts (p50.727). As the PTFE provides much greater X-ray attenuation than water there is less scope for modulation of the tube current when the PTFE inserts are scanned. As a result the slightly larger reduction in the reported tube current that was found when the water inserts were scanned results in a larger percentage change in noise relative to the scans with CARE Dose off. Figures 9 and 10 show that there were differences between the SNRs calculated for the scans with CARE Dose on and off. These figures also show that the SNR varied with position within the phantom. The highest values of image percentage noise and the lowest values of SNR were found in the central position which is as expected from photon path length and reconstruction theories. For PTFE the SNRs for the scans with CARE Dose on showed no distinct trend relative to the SNRs for the scans without CARE Dose. This is in contrast to the situation for water where the SNRs for the scans with CARE Dose on were lower than those for the scans with CARE Dose off for 80% of the total number of scans. This shows that there is a trend towards decreased SNR for water when CARE Dose is used. The larger set of SNR measurements showed a difference in the SNRs of approximately 10% for water whilst there was no difference for the PTFE measurements. This is attributable to the similar percentage change in the noise which was found (Table 2). Statistical analysis showed that there was no statistically significant difference in the SNRs for PTFE and water between the situations with and without CARE Dose (p50.197 for water, p50.764 for PTFE). Table 3 shows that, at the 95% level, there was also no significant change in the CT numbers for water and PTFE for the scans with and without CARE Dose. Since the SNR depends on both the signal and noise, neither of which showed a significant change at the 95% level, there was no associated significant change in the calculated SNRs for both water and PTFE. This does not provide an explanation for the users’ subjective opinions that the images acquired with CARE Dose, for imaging pelvic abscess, were unsatisfactory. When the SNR values for water are error corrected (mean value minus uncertainty), the average SNR for the scans with CARE Dose is only just above the detectability threshold of 5 as defined by Rose [12]. Water has an inherently low SNR relative to the PMMA background, but this is further reduced by 10% when CARE Dose is activated. The worse case SNR (i.e. the lowest value of SNR taking into account the calculated uncertainties) was below the threshold value of 5 for 25% of the measurements with CARE Dose off and for 40% of the measurements with CARE Dose on. Although these differences may not be statistically significant they may be detectable by the person viewing the image and are therefore important differences. The X-ray attenuation path in clinical scanning is nonhomogeneous and the human pelvis may have an even more asymmetric attenuation pattern than this phantom. This may introduce a larger modulation in the tube current which would affect the noise and serve to further worsen the SNR situation. This may therefore reduce the confidence with which the viewer of the image can detect tissues which have subtle differences in SNRs. This The British Journal of Radiology, January 2006

combination of the decrease in the water SNR and the non-homogeneous attenuation path may therefore explain why subjectively the images that were acquired with CARE Dose had been reported as unsatisfactory for pelvic abscess imaging. There were large uncertainties in the results of this study. However, the reductions in the SNRs that were found were repeatable over a large number of scans and are therefore considered to be a true representation of the performance of the system. The main explanation for the large uncertainties was that the ROIs that were used for the water and PTFE measurements were small – these were limited by the size of the inserts which were, in turn, limited by the construction of the phantom. If measurements were made too close to the edge of the insert then the mean CT number would have been skewed by the presence of any air around the insert or by the background material itself. It was not possible to make any changes to the phantom design. If it had been possible to use larger inserts (and therefore larger ROIs) it may have been possible to obtain results which were less error dominated. We would recommend that any future studies should consider using larger inserts and ROIs to improve the noise statistics and to ensure homogeneity in the measurements taken within the signal areas. However, it should be noted that at 12 mm in diameter the size of the water inserts were representative of abscesses which are found in the pelvis. Some differences were found between the results for the left and right lateral positions in the phantom, in terms of absorbed dose to air, noise and SNR for both inserts. Further tests showed that the central alignment laser was inaccurate by approximately 3 mm which resulted in a relative difference between the left and right measurement positions of around 6 mm and that the differences were not due to the performance of the CARE Dose system.

Conclusions The CARE Dose system on Siemens 4 slice CT scanners results in significant dose savings for scans of the pelvic region. This yielded a reduction of approximately 32% in the value of ‘‘CTDIvol’’ which agreed well with the 25% reduction in the displayed effective mAs. This implies that the reduction in the effective mAs value can be used as an approximate indicator of the true dose reduction. This reduction is a real, net dose saving as there was no statistically significant increase in the noise. There appears to be a trend towards decreased SNRs for both water and PTFE when CARE Dose was used although no significant differences were found at the 95% level. These changes in SNR were mostly due to changes in the image percentage noise values. The largest decreases in SNR were found for water and were as large as 14%. Since the water inserts were representative of low-density abscess this suggests that the use of CARE Dose may decrease the visibility of low-density structures relative to the background. Therefore using CARE Dose in situations where subtle differences in low CT number tissue pathology are of interest may not be advisable. 69

G R Iball, D S Brettle and A C Moore

Acknowledgments The authors wish to thank the following for their invaluable assistance in the work: Leeds Nuffield Hospital, especially Joanna Hartley for use of their scanner and for involvement in the measurement procedures; ImPACT, St George’s Hospital, London for loan of the phantom and general advice; Harrogate District Hospital, York District Hospital, UK CT Dose survey (2002) and Dr Paul Shrimpton of the National Radiological Protection Board (NRPB) for provision of protocol data; Siemens Medical (UK), Bracknell for protocol data.

References 1. Hart D, Wall BF. Radiation exposure of the UK population from medical and dental x-ray examinations (NRPB-W14). Chilton: National Radiological Protection Board, 2002. 2. Annals of the ICRP 26 (2). Radiological protection and safety in medicine. ICRP Publication 73, ICRP, 1996. 3. Kalender WA, Wolf H, Seuss C, Gies M, Greess H, Bautz WA. Dose reduction in CT by online tube current control: principles and validation on phantoms and cadavers. Eur Radiol 1999;9:323–328. 4. Gies M, Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys 1999;26:2235–47.

70

5. Kalender WA, Wolf H, Seuss C. Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements. Med Phys 1999;26:2248–53. 6. Greess H, Wolf H, Baum U, Lell M, Pirkl M, Kalender WA, et al. Dose reduction in computed tomography by attenuation based online modulation of tube current: evaluation of six anatomical regions. Eur Radiol 2000;10:391–4. 7. Greess H, Nomayr A, Wolf H, Baum U, Lell M, Bowing B, et al. Dose reduction in CT examination of children by an attenuation based online modulation of tube current (CARE Dose). Eur Radiol 2002;12:1571–6. 8. Mastora I, Remy-Jardin M, Seuss C, Scherf C, Guillot JP, Remy J. Dose reduction in spiral CT angiography of thoracic outlet syndrome by anatomically adapted tube current modulation. Eur Radiol 2001;11:590–6. 9. Jakobs TF, Becker CR, Ohnesorge B, Flohr T, Seuss C, Schoepf UJ, et al. Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 2002;12:1081–6. 10. Tack D, De Maertelaer V, Gevenois PA. Dose reduction in multidetector CT using attenuation based online tube current modulation. AJR Am J Roentgenol 2003;181:331–4. 11. Institute of Physics and Engineering in Medicine Report 32, Measurement of the Performance Characteristics of Diagnostic X-ray Systems used in Medicine Part III Computed Tomography X-ray Scanners, 2nd Edition, IPEM, 2003. 12. Rose A. Vision: human and electronic. New York: Plenum, 1974:21–3.

The British Journal of Radiology, January 2006