Comparison of size modulation and conventional

0 downloads 0 Views 2MB Size Report
May 5, 2016 - Conventional standard automated perimetry (SAP) has a constant stimulus ..... Although the k value used in Goldmann kinetic perimetry is a.
www.nature.com/scientificreports

OPEN

received: 19 August 2015 accepted: 20 April 2016 Published: 05 May 2016

Comparison of size modulation and conventional standard automated perimetry with the 24-2 test protocol in glaucoma patients Kazunori Hirasawa1, Nobuyuki Shoji1, Masayuki Kasahara2, Kazuhiro Matsumura2 & Kimiya Shimizu2 This prospective randomized study compared test results of size modulation standard automated perimetry (SM-SAP) performed with the Octopus 600 and conventional SAP (C-SAP) performed with the Humphrey Field Analyzer (HFA) in glaucoma patients. Eighty-eight eyes of 88 glaucoma patients underwent SM-SAP and C-SAP tests with the Octopus 600 24-2 Dynamic and HFA 24-2 SITA-Standard, respectively. Fovea threshold, mean defect, and square loss variance of SM-SAP were significantly correlated with the corresponding C-SAP indices (P