Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants Carl-F Wippermann, Dietmar Schranz, Ralf Huth, Fred Zepp, Hellmut Oelert, Bodo K Jiingst

was, however, reported to be unsatisfactory compared with the thermodilution method.78 The following factors are thought to contribute to the inaccuracy of the Doppler method: (a) Incorrect determination of vessel ison of the two techniques. For statistical diameter, which is subsequently squared evaluation one randomly assigned paired to calculate the cross sectional area of the vessel>--for example, measurement of the measurement of every patient was used. Setting-Paediatric intensive care unit aortic diameter as 11 mm instead of 10 mm in a neonate causes overestimation of cardiac in a university hospital. Patients-18 infants after open heart output by 21%. There are also appreciable surgery aged 4-25 months (weight 4- changes in cross sectional areas of aortic and pulmonary arteries during the cardiac cycle10 kg). Interventions-Cardiac output meas- for example 9% and 12% respectively in urements by dual beam Doppler and dogs.9 (b) Velocity measurements when the angle thermodilution techniques were performed within 10 minutes of each other of insonation (0) is > 150 can significantly and without knowledge of the results of underestimate cardiac output, especially if a the other methods. Multiple measure- "blinded" technique is used in which a small ments were performed on some patients angle 0 is assumed.' with a pharmacological or electro(c) Incorrect determination of the mean physiological intervention or with a min- velocity by measurement of the velocity at imum of six hours between each pair of only one point within the vessel. This velocity is used as the spatially averaged velocity measurements. Measurements and main results- assuming a flat velocity profile. It has been Three patients were excluded because of shown, however, that in both major arteries an inadequate Doppler signal or a sig- the velocity pattern changes.'°0 nificant residual shunt. Cardiac output The dual beam Doppler system is supposed measurements ranged from 0-4 to 2-2 to avoid many of the limitations of cardiac llmin for the thermodilution technique output determination by conventional and from 0O5 to 2-1 I/min for the dual Doppler echocardiography'2'14 and a dual beam Doppler technique. Agreement beam Doppler system is now available for between both methods was acceptable. infants and children. We have evaluated the The mean difference between the two accuracy of the dual beam device for measurmethods was 0-026 lImin with two stan- ing cardiac output in critically ill infants and dard deviations ranging from - 0-20 to children by comparing it with cardiac output 0-26 I/min. measured by thermodilution. Conclusion-The dual beam Doppler technique was shown to have promise for the non-invasive determination of Patients and methods cardiac output in critically ill infants. The study group consisted of 18 patients aged 4 to 25 months (mean 9 9 months) with a weight of 4 1 to 10 kg. The indication for cardiac Assessment of cardiac function is important in output measurement by thermodilution was the evaluation and treatment of critically ill management after open heart surgery. Surgery neonates, infants, and children. Cardiac out- included repair of an atrioventricular septal put is usually measured by thermodilution or defect in four patients and closure of a vendye dilution techniques. These techniques, tricular septal defect in 11 patients. We however, require insertion of a catheter into repaired tetralogy of Fallot, double outlet right the pulmonary artery.'3 They are complicated ventricle, and transposition of the great arteries in neonates and infants because the arteries in one patient each. The catheters were placed small size and are associated with low but in during operation. Multiple measurements measurable morbidity and mortality. Pulsed were made in some patients who had a pharDoppler measurement of cardiac output has macological or electrophysiological intervenbeen proposed as a non-invasive method of tion and in other patients who had a m iium measuring cardiac output in children and in of six hours between each pair of adults.' In critically ill patients technique measurements. Abstract

Objective-To compare cardiac output measurements in critically ill infants by the dual beam Doppler and thermodilution techniques. Design-Prospective direct compar-

University Children's Hospital, Mainz, Federal Republic of Germany C-F Wippermann D Schranz R Huth F Zepp B K Jingst Department of Cardiovascular Surgery, University of Mainz, Federal Republic of Germany H Oelert Correspondence to Dr Carl-F Wippermann, University Children's

Hospital, Langenbeckstrasse 1, D-6500 Mainz, Federal Republic of Germany. Accepted for publication 11 June 1991

Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants

181

DUAL BEAM DOPPLER TECHNIQUE: PRINCIPLE OF OPERATION

A commercially available dual beam Doppler system (Quantascope, Vital Science) with a 4 MHz two element annular array transducer (paediatric probe) was used. The "attenuation compensated flowmeter""5 used for this technique consists of a non-imaging, two parallel channel pulsed Doppler system with two concentric ultrasound beams and this approach has been described extensively.114 Briefly, the first beam is a uniform wide beam, with a sample volume that totally encompasses the vessel lumen (fig 1). The second, a narrow beam, is wholly within the flowing blood and its area at any depth is known exactly. Because of the size ofthe narrow beam, the diameter ofthe ascending aorta should be at least 10 mm. Mean velocity (V) is measured with the wide beam, which completely insonates the ascending aorta and thus determines the Doppler shift frequency of all erythrocytes passing through the aorta. Thus a constant velocity profile does not have to be assumed; V is then proportional to the mean Doppler shift frequency (Fd) divided by cos 6 of the angle of insonation: V oc Fd/cos 0 (1) The cross sectional area of the ascending aorta (A) is calculated from the intensity of the returned Doppler shift frequencies of the two beams. Because each erythrocyte contributes equally to this backscattered Doppler power, the Doppler power of the wide beam (Pw) is equal to the projected lumen area (Ap) multiplied by an unknown attenuation constant (Ka): Pw= Ap x Ka (2 A divided by the cos 0, Because AP is equal equation (2) can be written as: to

A x Ka w

(2

cos

So in equation (2) the Doppler po' wer of the narrow beam (P.) equals the known area of the narrow beam (A.) multiplied by the same attenuation constant Ka:

Pn= An

(44)

x Ka

A combination of equations (3) the cross sectional area A as:

anc1

(4) gives

Pw

A = A. x cos 0 x (5) the Cardiac output (CO) is producct of V, A, and heart rate (HR). Therefore comt)ination of equations (1) and (5) gives, cardiac output as: C

x

where C is

a

CO

=

An

x

Fd

x

Pn x HR

Figure I Schematic drawing showing the two beams of the dual beam Doppler in afront view of the heart and ascending aorta. Pn and Pw, returned Doppler power from the narrow and wide beams; Fd, mean Doppler shift.

DOPPLER CARDIAC OUTPUT MEASUREMENTS

The transducer was placed in the suprasternal notch and directed towards the ascending aorta. The transducer position and sample depth were optimised by the typical audio signal of aortic blood flow and by maximisation of the power return of the small beam. The power and velocity wave forms were evaluated with an online computer (Toshiba 3100). Dependent upon the heart rate of the patient, five to 15 beats were averaged and measurements of cardiac output, stroke volume, heart rate, and flow acceleration were obtained. The precision of cardiac output determination of the first 16 patients was limited to 0-1 1/min; with changed software it was improved to 0-001 1/min. Measurements were made without knowledge of the cardiac output by thermodilution. MEASUREMENT OF CARDIAC OUTPUT BY THERMODILUTION

(6)

precalibrated attenu;ation con-

stant.

The dual beam Doppler technicque determines Doppler shift frequency, Dop]pler power return of the two beams, and heart r-ate. Equation (6) is used to calculate cardiac (output (fig 1). Cardiac output measurement s by this device are independent of the angle of insonation 0 and a separately determined cross sectional area.

Cardiac output was measured by thermal dilution within 10 minutes of the Doppler determination of cardiac output without knowledge of those results and in an unchanged haemodynamic state. The study protocol consisted of three iced dextrose injections, which were averaged. Computation of cardiac output was done either with an Abbot Oximetrix 3 cardiac output computer or with the cardiac output module of the Mennen Medical Solo monitor system. The precision of cardiac output determination was limnited to 0-1 1/min.

Wippermann, Schranz, Huth, Zepp, Oelert, Jungst

182

A (n = 15)

STATISTICAL ANALYSIS

For the statistical evaluation of the agreement between the measurements of cardiac output by thermodilution and dual beam Doppler techniques only one randomly assigned paired measurement of each patient was used. The mean of each pair of measurements, together with the difference between each pair, were determined and these were plotted against each other. Agreement between the two methods was estimated by calculating the mean difference between the two methods and two standard deviations (SD) as suggested by Bland andAltman. 16 The difference between the two measurements ofeach pair was evaluated in absolute values and also as a percentage of the average cardiac output of each pair:

n.--,9f

2

., a _

E

01( 0

CL

.................

x 100

(CO thermodilution + CO Doppler) . 2

Reproducibility studies in adults showed that a difference of at least 15% (with three determinations for one measurement) between successive measurements is necessary to indicate a significant change in cardiac output.'7 For that reason percentage differences of > 15% between results from thermodilution and Doppler techniques were regarded as clinically meaningful. To illustrate the whole scatter of differences, all pairs of measurements were also evaluated; although it is not total, these pairs show some independence because of the changes that occur in the patient's clinical state between each pair of measurements.

lb

4-

2k 0._ o u

-0 4 I. 00

05

1.5

1.0

20

25

B (n = 96)

CO

4] r-

C-, a) c a)

a

0o;

-

2

a

U

*

*2-22 4a224 2-

0a)

U

a

2 225

*

o -0.: -0*4

a *

5 2-

=_* .2 _ _ _ _

0o0

% difference =

CO thermodilution - CO Doppler

0*4

0.5

1.5

1.0

__

20

25

Mean cardiac output (I/min) Figure 2 Differences in il/min between cardiac output determinations by thermodilution and Doppler plotted against the mean of the two determinations. (A) Only one randomly assigned paired measurement of every patient (n = 15) is shown. (B) All paired measurements (n = 96) are shown. Each square represents a measurement on one pair. Where two or more measurements fall on the same point they are indicated by the appropriate number. A (n = 15) 30 20 10 QQ+

_~~~~~~~*

10

-

0. -30 -20

-CO -30 0U5

U

1.0

15

20

Ii 2-

20

2z5

B (n = 96)

,In

'a) 1c0

Results Three of the 18 patients were excluded-two because of an inadequate Doppler signal. One of the two had a residual left ventricular outflow obstruction with high blood flow velocities that caused aliasing. The other had transposition of the great arteries. The third patient was excluded because ofa significant residual shunt, as shown by colour Doppler echocardiography and by an increase in oxygen saturation from the right atrium to the pulmonary artery. This would have caused an overestimation of cardiac output by the thermodilution method. We obtained 96 pairs of measurements from the remaining 15 patients. Cardiac output determinations ranged from 05 to 2-1 1/min for the Doppler technique and from 04 to 2-2 1/min for the thermodilution method. In the 15 pairs of measurements used for statistical evaluation, an acceptable agreement was found between the two methods, with a mean difference of 0-026 (2SD range -0-20026) 1/min (fig 2). The mean percentage difference between the methods was 1-9% (2SD range -20-5-24-2)% (fig 3). In one patient the percentage difference was > 15%. In four patients the percentage difference was borderline at about 15%. When all 96 pairs of measurements were used similar results were obtained (table, figs 2 and 3). In 13 pairs of measurements the percen-

30 1 a)-)2( C O~o

CE

DOCO

0) 0

2

20 10

-10 -

225

2~~-I

-

-20 -30hfi

0o0

2

*U 22 82 54wi 4 2 *223 2 4 2

-

05

5 2 .2_

.2

1.0

2

1.5

Mean cardiac output (I/min) Figure 3 Differences between cardiac output determinations by thermodilution and Doppler as percentages of the mean of the two determinations plotted against that mean. Explanation as forfig 2.

tage difference was > 15% and in another six this was borderline at about 15%.

Discussion In the group of critically ill children studied, cardiac output measurements by the dual beam Doppler and thermodilution methods showed acceptable agreement. This is the first study to evaluate the dual beam Doppler method in critically ill children. In adults as good, or better, results have been reported with the dual beam Doppler system in patients during cardiac catheterisation and in medical or surgical intensive care units.'2"'4 Two other studies in patients in intensive care found good correla-

Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants Agreement between dual beam Doppler and with thermodilution technique Thermodilution v Doppler One pair of

All pairs of

measurenents measurements (n = 96) per patient (n = 15) Mean difference

0-026 (0 23)

0-018 (0-25)

Mean percentage difference

1-88 (22-36)

0-65 (22 37)

(2SD) (l/min) (2SD) (%)

tions between cardiac output measured by the dual beam Doppler and thermodilution techniques. It was also noted that in some patients significant differences between the two methods could be shown by bias analysis.'8"9 In our study we found a considerable variation between results obtained by the two techniques: 95% of the percentage differences could be expected in the range -21% to 25%. In only one of the 15 paired measurements and in 13 of all paired measurements percentage differences > 15% were found and these were considered to be clinically meaningful. In a recent study that compared cardiac output measured by conventional Doppler echocardiography and by thermodilution in critically ill children a percentage difference > 15% between the two methods was noted in 45% of all measurements and a percentage difference > 25% was evident in a quarter of the measurements.7 Compared with these results the dual beam Doppler technique seems to give a more reliable measure of cardiac output in critically ill children. This is probably owing to the elimination of possible errors of the conventional Doppler method, as mentioned previously. The accuracy of the dual beam Doppler technique could probably be increased by averaging three measurements as in the thermodilution technique. In this technique the difference between two successive determinations necessary to show a significant change in cardiac output can be reduced from 26% to 15% if an average of three measurements is taken.'7 None the less the accuracy of the thermodilution method in humans is not known exactly and minor alterations in technique can cause overestimation of cardiac output by as much as 58% .20 21 Even under optimum conditions in vitro the standard error is seldom less than 10%.22 In a recent reproducibility study only 70% of the differences were within 20% of the mean value.23 It has been concluded that the overall biological error of any method of cardiac output measurement is 15-20%.24 Both Doppler methods, when used from the supersternal notch, calculate cardiac output without coronary blood flow. This explains, at least to some extent, the underestimation of Doppler cardiac output measurements compared with the thermodilution method in this study. Furthermore, the thermodilution method might have overestimated cardiac output because of small residual shunts that are often present immediately after patch closure of ventricular septal defects and which subsequently close spontaneously.25 For most of our patients colour Doppler echocardiography

183

was not available in the early postoperative period. Colour Doppler echocardiography carried out later showed a residual shunt only in the one patient mentioned earlier, who was excluded from the study population. The dual beam Doppler method is subject to several systematic errors. Unless the wide beam completely insonates the aorta cardiac output will be underestimated. Similarly the sample volume of the narrow beam has to be completely within the aorta, otherwise overestimation results. Furthermore, the wide beam may not reach the complete ascending aorta because of overlying lung tissue. On the other hand, the main or right pulmonary artery may also be partly insonated by the wide beam. Both of these factors will cause errors in calculated cardiac output. Because the Doppler power return of the two beams is used to calculate aortic cross sectional area, this calculation, as with conventional Doppler echocardiography, seems to be a source of possible major errors. The problem may be aggravated when the dual beam Doppler technique is used over the whole paediatric age group. Another Doppler probe with larger sized beams may be needed for older children who are still too small for the adult probe. Premature infants with low and very low birth weight need a probe with even smaller beams. We concluded that the dual beam Doppler technique is a simple and promising method to measure the ascending aortic blood flow noninvasively in critically ill infants. It seems especially useful in this age group, in which invasive cardiac output measurement is more difficult than in older children or the adult population. Further evaluation of this system including its use in older children is needed. We thank Paul Seifert, PhD, for his advice in the preparation of the paper. 1 Pollack MM, Reed TP, Holbrook PR, Fields AI. Bedside pulmonary artery catheterization in pediatrics. J Pediatr

1980;96:274-5.

2 Swedlow DB. Invasive assessment of the failing circulation. In: Swedlow DB, Raphaely RC, eds. Cardiovascular problems in pediatric critical care. New York: Churchill Livingston, 1986:112-49. 3 Tabata BK, Kirsch JR, Rogers MR. Diagnostic tests and technology for pediatric intensive care. In: Rogers MR, ed. Textbook of pediatric intensive care. Baltimore: Williams and Wilkins, 1987:1401-32. 4 Alverson DC, Eldridge M, Dillon T, Yabek SM, Berman W. Noninvasive pulsed Doppler determination of cardiac output in neonates and children. J Pediatr 1982;101:

46-50.

5 Goldberg SJ, Sahn DJ, Allen HD, Valdes-Cruz LM, Hoenecke H, Carnahan Y. Evaluation of pulmonary and systemic blood flow by two-dimensional Doppler echocardiography using Fast Fourrier transform spectral analysis. Am J Cardiol 1982;50:1394-9. 6 Huntsman LL, Steward DK, Barnes SR, Franklin SB, Colocousis JS, Hessel EA. Noninvasive Doppler Determination of cardiac output in man. Clinical validation.

Circulation 1983;67:593-602.

7 Notterman DA, Castello FV, Steinberg C, Greenwald BM, O'Loughlin JE, Gold JP. A comparison of thermodilution and pulsed Doppler cardiac output measurement in critically ill children. J Pediatr 1989;115:554-60. 8 Donovan KD, Dobb GJ, Newman MA, Hockings BES, Ireland M. Comparison of pulsed Doppler and thermodilution methods for measuring cardiac_output in critically ill patients. Crit Care Med 1987;15:853-7. 9 Loeber CP, Goldberg SJ, Marx GR, Carrier M, Emery RW. How much does aortic and pulmonary areary area vary during the cardiac cycle? Am Heart J 1987;113:95-100. 10 Redel DA. Color blood flow imaging of the heart. Berlin, Heidelberg, New York: Springer, 1989:31-42. 11 Jenni R, Vieli, A, Ruffinann K, Krayenb6hl HP, Anliker M. A comparison between single gate and multigate ultrasonic Doppler measurements for the assessment of

Wippermann, Schranz, Huth, Zepp, Oelert, Jungst

184

12

13

14

15 16 17

18

the velocity pattern in the human ascending aorta. Eur Heart J 1984;5:948-55. Evans JM, Skidmore R, Luckman NP, Wells PNT. A new approach to the noninvasive measurement of cardiac output using an annular array Doppler technique-I. Theoretical considerations and ultrasonic fields. Ultrasound Med Biol 1989;15:169-78. Evans JM, Skidmore R, Baker JD, Wells PNT. A new approach to the noninvasive measurement of cardiac output using an annular array Doppler technique-II. Practical implementation and results. Ultrasound Med Biol 1989;15:179-87. Looyenga DS, Liebson PR, Bone RC, Balk RA, Messer JV. Determination ofcardiac output in critically ill patients by dual Doppler echocardiography. J Am Coll Cardiol 1989; 13:340-7. Hottinger CF, Meindl JD. Blood flow measurement using the attenuation compensated volume flow meter. Ultrason Imaging 1979;1:1-15. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307-10. Stetz CW, Miller RG, Kelly GE, Raffin TA. Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 1982;126:1001-4. Niclou R, Teague SM, Lee R. Clinical evaluation of a

19 20 21 22

23

24

diameter sensing Doppler cardiac output meter. Crit Care Med 1990;18:428-32. DH, Mahutte CK. Two-beam cardiac output measurement: reproducibility and agreement with thermodilution. Crit Care Med 1990;18:432-7. Marushak GF, Potter AM, Schauble JF, Rogers MC. Overestimation of pediatric cardiac output by thermal indicator loss. Circulation 1982;65:380-3. Schuster AH, Nanda NC. Doppler echocardiographic measurement of cardiac output: comparison with a non golden standard. Am J Cardiol 1984;53:257-9. MackenzieJD, Haitres NE, Rawles JM. Method of assessing the reproducibiity of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers. Br Heart J 1986;55:14-24. Russell AE, Smith SA, West MJ, Aylward PE, McRitchie RJ, Hassam RM, et al. Automated non-invasive measurement of cardiac output by the carbon dioxide rebreathing method: comparisons with dye dilution and thermodilution. Br Heart J 1990;63:195-9. Levitt JM, Replogle RL. Thermodilution cardiac output: a critical analysis and review of the literature. J Surg Res

Wong

1979;27:392-404.

25 Stevenson JG, Kawabori I, Stamm SJ, Bailey WW, Hall DG, Mansfield PB, et al. Pulsed Doppler echocardiographic evaluation of ventricular septal defect patches. Circulation 1984;70(suppl I):38-46.