Domestic Development of Single-Photon Emission ...

2 downloads 0 Views 542KB Size Report
Jan 26, 2017 - Yang Shuang, Dou Xiu-Ming, Yu Ying et al. Single-Photon Emission from a Single InAs Quantum Dot. Dou Xiu-Ming, Sun Bao-Quan, Huang ...
Home

Search

Collections

Journals

About

Contact us

My IOPscience

Domestic Development of Single-Photon Emission Computed Tomography (SPECT) Unit with Detector based on Silicon Photomultipliers

This content has been downloaded from IOPscience. Please scroll down to see the full text. 2017 J. Phys.: Conf. Ser. 784 012017 (http://iopscience.iop.org/1742-6596/784/1/012017) View the table of contents for this issue, or go to the journal homepage for more Download details: IP Address: 184.174.98.11 This content was downloaded on 26/01/2017 at 16:39 Please note that terms and conditions apply.

You may also be interested in: Single-Photon Emission from GaAs Quantum Dots Embedded in Nanowires Yang Shuang, Dou Xiu-Ming, Yu Ying et al. Single-Photon Emission from a Single InAs Quantum Dot Dou Xiu-Ming, Sun Bao-Quan, Huang She-Song et al. AN INTEGRATED-CIRCUIT PULSE AMPLIFIER FOR USE WITH PHOTOMULTIPLIERS D. L. DuPuy Multichannel analog front-end and analog-to-digital converter ICs for silicon photomultipliers Y I Bocharov and V A Butuzov Application of ultrashort laser pulses for timing characterization of silicon photomultipliers E V Popova, P Zh Buzhan, A A Stifutkin et al. Quantum interference between Raman scattering and single-photon absorption Li Yan-Rong and Wang Yun-Ping Performance of SiPMT array readout for fast time-of-flight detectors M Bonesini, R Bertoni, A de Bari et al. A solvable operator motion equation of an excited-atom coupled leaky cavity in single-photon generation Chai Jin-Hua and Han Zheng-Fu Single-photon emission correlated to double-electron capture by bare ions: background processes T Elkafrawy, A Warczak, A Simon et al.

International Symposium Physics, Engineering and Technologies for Bio-Medicine IOP Publishing IOP Conf. Series: Journal of Physics: Conf. Series 784 (2017) 012017 doi:10.1088/1742-6596/784/1/012017

International Conference on Recent Trends in Physics 2016 (ICRTP2016) IOP Publishing Journal of Physics: Conference Series 755 (2016) 011001 doi:10.1088/1742-6596/755/1/011001

Domestic Development of Single-Photon Emission Computed Tomography (SPECT) Unit with Detector based on Silicon Photomultipliers S Grishakov¹, O Ryzhikova², V Sergienko³, A Ansheles4 and S Novikov5 ¹ UNIX medical company, Moscow, Russia ² Russian Cardiology Research and Production Complex, Moscow, Russia ³ Russian Cardiology Research and Production Complex, Moscow, Russia 4 Russian Cardiology Research and Production Complex, Moscow, Russia 5 Institute of oncology, St.Peterburg, Russia

E-mail: [email protected] Abstract. The idea of creating a single-photon emission computed tomography unit with solid-state photomultipliers is not new [1], as the problems of analog-to-digital conversion with a lot of noise and a wide range of values of intrinsic spatial resolution of the detector in a center and relevant fields of view could not be solved by means of gamma-camera detector architectures based on vacuum photomultipliers. This paper offers a new SPECT imaging solution that is free from these problems.

1. Introduction Radionuclide diagnostics is a technique that allows evaluating structure-functional state of organs and tissues of human body by means of different diagnostic radiopharmaceuticals administered in the body to acquire imaging information about the state of cellular systems under examination. Radiopharmaceuticals are specific chemical or biochemical compounds labeled with gammaemitting radionuclides with a short half-life. Gamma radiation emitted from the body of a patient is recorded by a gamma camera detector and the received information is converted into a functional image of the examined organ after computer processing. Spatiotemporal pattern of the radiopharmaceutical distribution provides the information about shape, size and location of the organ, as well as about presence of lesions in it. Radionuclide diagnostics is widely used in oncology (brain tumors, scintigraphy of malignant breast tumors, bone and liver scintigraphy), endocrinology, cardiology, urology and nephrology. 2. From prototype to industrial design Nowadays, the most common gamma camera units are based on vacuum photomultipliers. Known user problems are defined by specifications of gamma camera detectors [2], this can be described by two values having determining competitive advantage at the market. These are the values of the spatial resolution of the detector in the center and in the relevant field of view at the level of 0.5 (FWHM) and the value of energy resolution of the unit for 99mTc. The problem of detector energy resolution values for 99mTc gets prominent coverage in literature as well as in this article. The authors focus on devices

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. 1 Published under licence by IOP Publishing Ltd

International Symposium Physics, Engineering and Technologies for Bio-Medicine IOP Publishing IOP Conf. Series: Journal of Physics: Conf. Series 784 (2017) 012017 doi:10.1088/1742-6596/784/1/012017

electronic component which determines detector energy resolution values. For silicon photomultiplier detector, channels number defines this value. Channel is a complex of the silicon photomultiplier (detector pixels), buffers and comparators with external control, analog-digital, time-to-digital converters and high-speed data bus. The creation of a gamma camera detector with the field of view of 400 x 540 mm2, based on silicon photomultiplier, requires minimum 13824 pixels. This detector architecture allows obtaining the values of the spatial resolution of the detector in the center and in the relevant field of view at the level of 0.5 (FWHM) with minimal differences. Practically, the absolute values equality could not be obtained [3] because the measured modulation transfer function is unimodal and similar to a Gaussian function [4]. The values of 2.9 mm at the level of 0.5 FWHM and of 5.3 mm at the level of 0.1 FWHM were obtained. This fact is confirmed by the advance advertising of the new unit by General Electric [5]. The unit parameters of the spatial resolution are 2.5 mm and 3.8 mm respectively. The value of the internal energy resolution for 99mTc [6] is determined by a manufacturer during the unit acceptance tests and depends on the specified detector architecture. In our design the detector is made of large single-crystal CsI(Tl) pixels, and silicon photomultipliers are joined by optical glue. Such architecture and recovery algorithm for gamma event allow reaching internal energy resolution of 9.2% for 57Co [7]. According to authors [8], the energy resolution of the detector is inversely proportional to the square root of the energy of gamma quanta of the radioactive isotope used in the measurement. Consequently, the recalculation of the internal energy resolution for 99mTc gives the value of 8.5%. Obviously, the cost of single-photon emission computed tomography units with solid-state pixels is higher than the cost of units with detectors based on the vacuum photomultipliers. This situation will not change until the market is filled with modern analogues. For this purpose, the price of components of such units should fall for a manufacturer. For detectors on silicon photomultipliers, this is decisive for entering the competitive market. Alternative way, when a manufacturer introduces new technology, can be applied in the modernization of the existing technical base. GE used this option for the introduction of CZT technology. Other manufacturers will do the same. The time will show if we should look for our own way to solve the problem or we should rely upon the experience of the colleagues. 3. Conclusions Nowadays there is a tendency towards the use of modern materials in gamma-camera detectors and SPECT. The SPECT imaging unit was developed in Russia. Its parameters can compete with world latest developments. The standard [4] underlines that its application to the detectors with the architecture of many pixels is incorrect. So we need new applications to the standards for a new generation of gamma cameras, and a direct comparison of the specifications of gamma camera based on the vacuum photomultipliers and silicon photomultipliers is also incorrect. However, the market has its own rules, and a buyer will choose a cheaper unit capable to satisfy minimal necessary requirements of a user. References [1] Narkevich B, Physical and technical support of nuclear medicine: current state and development prospects, , Radiation oncology and nuclear medicine, No1, 2012. [2] Arsvold D, Vernik M. Emission Tomography: Fundamentals of PET and SPECT. Technosphera, Moscow, 2009. [3] UNIX Medicl Company. Test protocol No. 01-1 of 12 August 2015 for production prototype of gamma-camera No.13411.1008799.13.108 The Ministry of Industry and Trade of the Russian Federation, 2015. [4] NEMA NU 1-2012, Performance Measurements of Gamma Cameras, National Electrical Manufacturers Association, P. 8,9, 2013. [5] http://www3.gehealthcare.com/en/products/categories/nuclear_medicine/spect_and_spect-

2

International Symposium Physics, Engineering and Technologies for Bio-Medicine IOP Publishing IOP Conf. Series: Journal of Physics: Conf. Series 784 (2017) 012017 doi:10.1088/1742-6596/784/1/012017

[6] [7] [8]

ct/discovery_nm-ct_670_czt NEMA NU 1-2012, Performance Measurements of Gamma Cameras, National Electrical Manufacturers Association, pp. 11-13, 2013. GOST R IEC 60789-2010, Characteristics and test conditions of radionuclide imaging devices. Anger type gamma cameras. Standartinform, Table 1, p. 3, 2012. Vartanov N, Samoylov P, Applied scintillation gamma-spectroscopy, Atomizdat, Moscow, 1975.

3