Tribol Lett (2011) 44:387–397 DOI 10.1007/s11249-011-9856-9

METHODS PAPER

Friction Measurements on Contact Lenses in Their Operating Environment M. Roba • E. G. Duncan • G. A. Hill N. D. Spencer • S. G. P. Tosatti



Received: 14 April 2011 / Accepted: 5 September 2011 / Published online: 17 September 2011  Springer Science+Business Media, LLC 2011

Abstract An important issue concerning the use of soft contact lenses is comfort, which, among other factors, has been related to the level of friction between the anterior side of the lens and the inner eyelid. Although several studies have been carried out to investigate the frictional properties of contact lenses, these have not taken the physiological environment of the eye into account. In use, lenses are in contact with proteins present in tears, with corneal cells and with the palpebral conjunctiva (clear membrane on inner eyelid). The focus of this study was to establish a biologically relevant measurement protocol for the investigation of friction of contact lenses that would mimic the eye’s physiological environment. By optimizing parameters such as the composition of the friction counter surface, the lubricant solution, the normal load and the velocity, an ideal protocol and setup for microtribological testing could be established and used to perform a comparative study of various commercially available soft contact lenses.

M. Roba  E. G. Duncan  G. A. Hill  S. G. P. Tosatti (&) SuSoS AG, Lagerstrasse 14, CH-8006 Duebendorf, Switzerland e-mail: [email protected] G. A. Hill  N. D. Spencer Department of Materials, Laboratory for Surface Science and Technology, ETH Zurich, Wolfgang-Pauli-Strasse 10, CH-8093 Zurich, Switzerland G. A. Hill VISTAKON, A Division of Johnson & Johnson Vision Care Inc, Jacksonville, FL, USA G. A. Hill GHBiomaterials, LLC 1918 Hickory Lane, Atlantic Beach, FL 32233, USA

Keywords Contact lens  Microtribometer  Mucin  Friction  Hydrogel

1 Introduction The use of soft contact lenses is extremely widespread [1]. However, issues concerning end-of-day comfort still exist [2]. Discomfort caused by contact lenses has been related to several factors such as dryness, protein adsorption, physiological factors and friction occurring during the blinking process, especially between the anterior side of the lens and the inner eyelid [3–8]. Clinical tests have been performed to understand and improve comfort, and have centered on the trial of different lens materials and ophthalmic solutions [9–13]. The dryness issue has been investigated by studies of the incorporation of wetting agents into contact lenses [8] and on the effect of water content of the lens [14]. A few studies have been carried out to investigate friction of the lens [4–7]. Blinking is the primary physiological contributor to the forces exerted on contact lenses and their consequent motion. Typical values of contact pressure are in the range of 3–5 kPa, sliding speed being around 12 cm/s [15, 16]. The friction studies carried out to date have been based on these reported values, despite the practical difficulties in achieving them because of resolution limits of the microtribometers used. However, the physiological environment has never been taken into account. Rennie et al. [4] investigated the friction between glass and a contact lens by positioning the lens taken from borate buffer on a special holder, without additional lubricant being added. Nairn and Jiang [5] looked at friction, for both the anterior and posterior side of contact lenses, against polymethylmethacrylate (pMMA) and polyhydroxyethylmethacrylate (pHEMA) counter

123

388

surfaces, using ophthalmic solutions as lubricants. Dunn et al. [7] carried out studies on friction between the anterior side of a contact lens and corneal epithelial cells, while Dong and Haugstad [17] looked at friction and adhesion of polyvinylalcohol (PVA) contact lenses by scanning probe microscopy (SPM). Lydon et al. [18] studied the coefficient of friction (CoF) for contact lens versus glass and polyethylene (PE) and, lastly, Ngai et al. [6] investigated friction for a contact lens versus glass system in the presence of saline solution. Interestingly, they also considered the possible effect of protein adsorption from the tear film by incubating some of their lenses in a lysozyme and albumin solution. Apart from the work performed by Ngai et al., these studies generally did not take physiological aspects, such as the eyelid counter surface and the presence of a tear film, into account. Under the physiological conditions on the eye surface, the contact lens is in contact with proteins present in tears and the anterior side of the palpebral conjunctiva (inner eyelid) [7]. This study has shown that measuring contact lens frictional properties without taking the physiological environment into account can lead to misleading results. It is, therefore, important to perform such measurements under biologically relevant conditions. Friction tests on a glass versus contact lens system were performed by means of a microtribometer. The measurement protocol was optimized by varying the functionalization of the glass counter surface, with the aim of mimicking the on-eye environment. A comparative study was carried out by measuring the coefficient of friction of several commercially available contact lenses using the newly developed measurement protocol.

Tribol Lett (2011) 44:387–397

Fig. 1 Silicone-rubber cover (left hand side) and PMMA ring (right hand side). Silicone cover was cast to perfectly match the geometry of the Teflon chamber, rounded holder and contact lens system. A hole was made on the top part of the silicone cover to allow the contact lens anterior surface to be exhibited

2 Materials and Methods 2.1 Instrument and Setup Friction tests were performed with a microtribometer (Basalt Must, Tetra, Germany). Cantilevers (Tetra, Germany) with different ranges of spring stiffness (N/m) were used: kn = 23, kt = 23, ±10% and kn = 15, kt = 15, ±10% (kn is the normal force spring constant and kt is the tangential force spring constant). The contact lens was placed inside a Teflon chamber on top of a sand-blasted rounded plastic holder (cyclo olefin polymer, Johnson & Johnson Vision Care Inc., USA), matching the internal radius of curvature of the lens, and was held in position by a cast silicone-rubber cover (polyvinylsiloxane, Provil Novo, Germany) and plastic ring (poly(methyl methacrylate), PMMA) (Figs. 1 and 2). Silicone cover and PMMA ring were screwed to the Teflon chamber by two screws placed at 180 to one another. The anterior surface of the

123

Fig. 2 Experimental setup for tribological tests: The contact lens is placed on a rounded holder inside a Teflon chamber. The green silicone-rubber cover and PMMA ring that hold the lens in position are screwed into the Teflon chamber

lens was facing upward. The counter surface consisted of a functionalized 5-mm diameter glass disk (cover glass, Thermo Scientific, Germany). Functionalization protocols are described in the following. A 6-mm long glass rod was glued onto the tip of the tribometer cantilever. In turn, the functionalized glass disk was glued to the glass rod, being the latter as centered as possible (Fig. 3). Gluing was

Tribol Lett (2011) 44:387–397

389

covered with lubricant solution during the friction test. Figure 5 shows an example of contact lens and counter surface mounted in the microtribometer. Contact area and pressure between the flat glass counter surface and the soft contact lens (Table 1) were estimated using the Hertzian contact model as described by Chaudhri and Yoffe [19]. They related the radius of the contact area ah to the mechanical properties of the materials by 

Fig. 3 Example of cantilever modified for friction experiment: The glass rod is glued to the cantilever tip; second, a glass disk is glued on the other end of the glass rod

a3h ¼ RP½ð1  m21 Þ=ðE1 Þ þ ð1  m22 Þ=ðE2 Þ



ð1Þ

R being the radius of the contact lens, P the load, mi Poisson’s ratio and Ei Youngs’s modulus. French reported the Young’s modulus for a series of contact lenses materials [20]. For the purposes of this publication, values of contact area and pressure were estimated for the contact lens materials with the highest and lowest moduli. These materials are: ACUVUE NIGHT & DAY

Fig. 4 Measurement setup: The functionalized glass disk, attached to the cantilever through a glass rod, is gently lowered until contact with the lens occurs. Arrows in the drawing indicate normal force and lateral force directions

performed using a cyanoacrylate-based glue (UHU GmbH & Co. KG, Germany). The cantilever was then mounted in the tribometer. Just before the test, the cantilever was slowly lowered so that the glass disk approaches the center of the contact lens surface (determined by eye) (Fig. 4). The contact lens was

0.3 MPa 

1.5 MPa

ACUVUE brand contact lens is a cast-molded polyHEMA contact lens manufactured by VISTAKON, a division of Johnson & Johnson Vision Care Inc., Jacksonville, FL, USA. The water content is around 58%. Focus Night & Day is a cast-molded silicone hydrogel contact lens manufactured by CIBA VISION. It has a water content of about 24% and has a plasma-treated surface. Poisson’s ratio has not been reported for these materials. A value of 0.3 was used in the calculations. The values for the glass counter surface are not important for this analysis. The Young’s modulus is much greater than that of the contact lenses; therefore, 1/E becomes negligible in Eq. 1. 2.1.1 Sample Preparation Contact lenses were removed from their packaging immediately before the test and installed on the sample holder, which had previously been cleaned with water and

Table 1 Contact area and contact pressure estimated values for ACUVUE and NIGHT & DAY pressing against a flat glass surface Normal force (mN)

Contact area (mm2) ACUVUE



Contact pressure (kPa) NIGHT & DAY



ACUVUE

NIGHT & DAY

2.50E - 01 5.00E - 01

0.093 0.15

0.032 0.057

2.7 3.4

7.8 9.5

1.00E ? 00

0.24

0.081

4.2

12.4

2.00E ? 00

0.37

0.13

5.4

15.6

3.00E ? 00

0.49

0.17

6.1

17.9

4.00E ? 00

0.59

0.20

6.7

19.7

123

390

Tribol Lett (2011) 44:387–397 Table 2 Advancing contact angle measurements for treated glass surfaces and ellipsometry thickness measurements for SiO2 model surfaces Treatment

Contact angle ()

Adlayer thickness (nm)

Plasma

\10

n/a (substrate base)

Hexamethyldisilazane

[75

0.25 (±0.04)

Mucin

\45

1.2 (±0.1)

Adlayer thickness measurements were performed on two separate occasions with a total of six measurements

Fig. 5 Contact lens and cantilever mounted in the microtribometer. The cantilever is slowly lowered until the glass disk is in contact with the contact lens

detergent (1:1) (hydrochloric acid 300 mmol/L, detergent 1%, Cobas Integra, Roche, Switzerland) and wetted with lubricant solution. Excess storage solution (from lens packet) trapped between the lens and the holder was avoided by gently tapping the edge of the lens on a clean surface. The contact lens was immediately covered with lubricant solution to avoid drying. 2.1.2 Counter Surface Preparation Before fixation onto the cantilevers, glass disks were oxygen plasma cleaned for 2 min (Diener Electronic, Germany) and hydrophobized with hexamethyldisilazane (Alfa Aesar, Germany) from the gas phase for 30 min in a vacuum desiccator. To assess successful hydrophobization, the dynamic water contact angle ([75) was measured on hydrophobized glass model surfaces. Contact angle (advancing (wCAadvancing) measurements were performed with a Kru¨ss contact angle–measuring system (G2/G40 2.05-D, Kru¨ss GmbH, Germany) with a drop speed of 15 lL/min. A movie with 100 images was recorded for the advancing contact angle and the analysis was carried out by means of the tangent method 2 routine of the Kru¨ss Drop-Shape Analysis program (DSA version 1.80.0.2 for Windows 9x/NT4/2000, 1997-2002 KRUESS). Hydrophobized glass disks were mounted on the cantilever as previously described. Immediately before the friction tests, the glass disk mounted on the cantilever was incubated in a 1 mg/mL mucin solution

123

(from bovine submaxillary glands type I–S, Sigma-Aldrich, Germany) in HEPES 1 (10-mM N-2-hydroxyethylpiperazineTM N-2 ethanesulfonic acid, pH 7.4, BDH , UK) for 30 min and rinsed with ultra-pure water to remove any non-adsorbed species. Additionally, hexamethyldisilazane and mucin layer thickness in the dry state were evaluated by ellipsometry (J.A. Woollam Co., Inc., Ellipsometry Solutions, USA) on oxidized silicon wafer model surfaces. The surface modification protocol applied to the silicon wafers was the same as for glass model surfaces, with additional initial pre-cleaning with toluene and isopropanol. Adlayer thickness measurements were conducted under ambient conditions at an angle of incidence of 70 with respect to the surface normal, averaging 50 measurements at each point. The spectral range was 370–995 nm. Data were fitted with the WVASE32 analysis software using a three-layer model (Si/SiO2/Cauchy), where Si was assumed to be constant for all wafers (1 mm). The SiO2 layer was fitted with the SiO2 model before hydrophobization with silanes and mucin adsorption; the silane and mucin layers were fitted using the Cauchy (organic) layer model. The dry film thickness of the silane layer and of the mucin layer was assumed to have a refractive index of 1.45. Characterization results are shown in Table 2. To mimic the surface of the eyelid, glass disks with different functionalizations were evaluated: oxygen plasma cleaning (hydrophilic), silanization (hydrophobic) and protein or polymer attachment. Proteins used were fibrinogen (from bovine plasma, fraction I type I–S, Sigma-Aldrich, Germany) and lysozyme (AppliChem, Germany), whereas poly(L-lysine)-graft-poly(ethylene glycol) (PLL(20)-g[3.5]PEG(2), SuSoS AG, Switzerland) was used for polymer coating. Plasma cleaning and silanization were carried out as described above. Fibrinogen and lysozyme coatings were achieved by immersing plasma-cleaned glass disks, already glued onto the cantilevers, in 1.5 mg/mL fibrinogen in phosphate-buffered saline (PBS) or 1 mg/mL lysozyme in HEPES 1 for 30 min. Finally, PLL-g-PEG functionalization was performed by immersing plasma-cleaned cantilevermounted glass disks into 0.1 mg/mL PLL-g-PEG solution in HEPES 1 for 30 min. After incubation, the disks were rinsed with ultra-pure water and dried with nitrogen. Results are discussed in Sect. 3.

Tribol Lett (2011) 44:387–397

391

2.1.3 Lubricant Solution

3 Results

A 0.9% sodium chloride solution with a borate buffer (pH * 7.4) was used as a control (packing solution [PS] supplied by Johnson & Johnson Vision Care Inc. [JJVCI]). As a lubricant solution, a tear-mimicking solution containing PS plus serum (Precinorm U, Roche, Germany) (20:1) plus lysozyme (5 mg/mL) was used to perform comparative friction tests described in Sect. 3.

3.1 Optimization Measurements

2.1.4 Experimental Conditions All experiments for the optimization step were carried out TM using 1•DAY ACUVUE and 1•DAY ACUVUE MOIST brand contact lenses. In particular, the influence of sliding cycles and velocity was determined using tear-mimicking solution as a lubricant and mucin-coated silanized glass disks as counter surfaces. The effect of glass disk functionalization to mimic the eyelid was evaluated using packing solution as lubricant. The comparative study was performed when a model to mimic the natural eye environment was established: All contact lenses were tested in tear-mimicking solution against mucin-coated silanized glass disks. 2.1.5 Measurement Program The measurement program used for the optimization step and the comparative friction tests involved applying three sets of seven normal forces varying from 0.25 to 5 mN. Two cycles, backward and forward, were recorded for each target normal force value. Only data points obtained from the second cycle were recorded and subsequently analyzed. Ageing was simulated by performing 50 additional cycles at 2-mN normal force between each set of seven normal forces. In detail, the measurement protocol consisted of the following steps: measurement of coefficient of friction (CoF) with seven normal forces, 50 ageing cycles, measurement of CoF with seven normal forces, 50 ageing cycles to give a total of 100 ageing cycles and again a final CoF determination. The measured stroke length was 1 mm and the sliding speed was 0.1 mm/s. During optimization of the measurement protocol, 1 and 10 mm/s sliding speeds were also investigated. For data processing, lateral and experimentally determined normal force values were calculated by averaging 20 data points at around 0.5-mm distance, and then plotted in a graph. Trace and retrace curves obtained in this way were averaged and a coefficient of friction was determined from the slope. 2.1.6 Contact Lenses The lenses used in this study are listed in Table 3. The lenses were supplied by JJVCI.

3.1.1 Effect of Number of Sliding Cycles Preliminary tests run on 1•DAY ACUVUE and 1•DAY TM ACUVUE MOIST brand contact lenses to investigate the influence of ageing on friction revealed that CoF at 100 cycles at a normal load of 2 mN (maximum ageing) is equal to or not significantly different than CoF at 0 cycles (no ageing) (Fig. 6), indicating that there is no change in the coefficient of friction for either lens type within the limits of this experimental protocol. 1•DAY ACUVUE TM MOIST contact lenses exhibited significantly lower CoF than 1•DAY ACUVUE contact lenses. Significant differences in the results were evaluated by means of statistical analysis (p value \0.05). 3.1.2 Effect of Sliding Velocity CoF was shown to significantly increase (p value \0.05) with sliding velocity for both 1•DAY ACUVUE and TM 1•DAY ACUVUE MOIST brand contact lenses (Fig. 7). Above 1 mm/s, distinguishing between the two lenses investigated was difficult because of high and overlapping standard deviations. 3.1.3 Choice of Counter Surface The CoF was measured for different counter surfaces in packing solution. Results show that the lowest CoF was obtained for PLL-g-PEG-coated glass disks, whereas higher CoF values were obtained for hydrophilic and lysozyme-coated glass disks (Fig. 8). In the case of hydrophilic glass disks, CoF for 1•DAY ACUVUE TM MOIST was significantly higher than for 1•DAY ACUVUE. In all other cases where there is a significant TM difference, 1•DAY ACUVUE MOIST contact lenses exhibited lower CoF than 1•DAY ACUVUE. No post measurement testing was performed on the mucin-coated counter surface. As the coefficient of friction did not change between the first ramp (0 cycles) and the final ramp (after 100 cycles), the lens surface and counter surface were deemed unchanged during the experiment. 3.2 Comparative Study Various commercially available daily disposable and reusable contact lenses were tested using the optimized protocol, tear-mimicking solution as lubricant and mucincoated silanized glass as counter surface. Friction

123

392

Tribol Lett (2011) 44:387–397

Table 3 Lenses used Daily disposable lenses Name of contact lens TM

Manufacture

Silicone hydrogel?

Surface treatment?

Sauflon

No*

Yes

No

Johnson and Johnson Vision Care, Inc.

Yes

No

No

(narafilcon B)

Johnson and Johnson Vision Care, Inc.

Yes

Yes

No

(narafilcon A)

Johnson and Johnson Vision Care, Inc.

Yes

Yes

No

1•DAY ACUVUE

Johnson and Johnson Vision Care, Inc.

No

No

No

Focus DAILIES TM All Day Comfort

CibaVision

Proclear 1 day

CooperCooperVision

CLARITI

1 day

Contains PVP?

1•DAY ACUVUE MOIST 1•DAY ACUVUE TruEye

TM

1•DAY ACUVUE TruEye

TM

ClearSight

TM

1 Day

DAILIES AquaComfort Plus

No

No

No

No

No

No

CooperVision

No

No

No

CibaVision

No

No

No

Reuseable lenses AVAIRA

CooperVision

No*

Yes

No

ACUVUE OASYS

Johnson and Johnson Vision Care, Inc.

Yes

Yes

No

TM

CLARITI

Sauflon

No*

Yes

No

ACUVUE ADVANCE PLUS

Johnson and Johnson Vision Care, Inc.

Yes

Yes

No

ACUVUE ADVANCE

Johnson and Johnson Vision Care, Inc.

Yes

Yes

No

Biofinity

CooperVision

No*

Yes

No

ACUVUE

Johnson and Johnson Vision Care, Inc.

No

No

No

[AIR OPTIX] NIGHT & DAY AQUA

CibaVision

No

Yes

Yes (methane air plasma)

PremiO AIR OPTIX AQUA

Menicon CibaVision

No No

No Yes

Yes Yes (methane air plasma)

CibaVision

No

Yes

Yes (methane air plasma)

CibaVision

No

Yes

Yes (methane air plasma)

BAUSCH?LOMB

No

Yes

Yes (oxygen plasma)

TM

AIR OPTIX (formerly O2 OPTIX ) NIGHT & DAY PureVision TM

Optima



38

BAUSCH?LOMB

No

No

No

SOFLENS

BAUSCH?LOMB

No

No

No

* Indicates no PVP is reported in formulation but maybe formed in situ

coefficient was evaluated before (0 cycles) and after ageing (50 and 100 cycles) (Tables 4 and 5; Figs. 9 and 10).

4 Discussion The goal in the development of a model is to be able to use the in vitro data to predict in vivo performance. To correlate clinical contact lens performance with the model results and to evaluate potential sources of comfort-driven performance, possible lens discontinuation and adverse effects to the coefficient of friction, it is desirable to

123

measure the CoF of contact lenses in the most biologically relevant conditions. In the physiological environment, contact lenses are in contact with proteins present in tears, with corneal cells (on the posterior side) and with the palpebral conjunctiva (on the anterior side). Most studies carried out so far on the tribology of contact lenses have not taken such physiological conditions into account [4–6, 18]. The coefficient of friction is a property of the sliding pair; therefore, the determination of the coefficient of friction using different counter surfaces would be expected to produce different results. When comparing the results found in this study

Tribol Lett (2011) 44:387–397

Fig. 6 Effect of number of sliding cycles (ageing) on friction using TM 1•DAY ACUVUE and 1•DAY ACUVUE MOIST brand contact lenses versus hydrophobized mucin-coated glass disks. Tear-mimicking solution was used as lubricant

Fig. 7 Graph showing the effect of sliding velocity on friction for TM 1•DAY ACUVUE and 1•DAY ACUVUE MOIST brand contact lenses versus hydrophobized mucin-coated glass disks in tearmimicking solution. Error bars are standard deviation of three repeat experiments

with those from previous studies, the following observations can be made. Rennie et al. [4] evaluated Etafilcon A using an untreated spherical glass counter surface at normal forces of 3–20 mN. They report that the coefficient of friction obeys a sliding speed–dependent power law. Using their data, a coefficient of friction of about 0.15 would be predicted for Etafilcon A. Our results, on the other hand, show a coefficient of friction of 0.02–0.09 that was invariant with normal force. Nairn and Jiang [5] evaluated the coefficient of friction for SeeQuence (Polymacon) brand contact lenses using a pin-on-disk tribometer. They report coefficients of friction of 0.05–0.3 in saline. Our TM results for Optima 38, a Polymacon lens produced by the same manufacture, were 0.55. Finally, Ngai et al. [6] reported the coefficients of friction for Lotrafilcon A and Polymacon A to be 0.27 (calculated from normal and tangential forces in their study). Our results were 0.38 and

393

Fig. 8 Mimicking the eyelid: CoF at 100 cycles for various functionalized glass counter surfaces measured in packing solution. Sliding speed was 0.1 mm/s and normal forces ranged from 0.25 to 5 mN. Error bars are standard deviation of three repeat experiments

0.55, respectively. These variations show the need for a standardized method for the measurement of the coefficient of friction for contact lenses. Protein adsorption on contact lenses was investigated by Garrett and Milthorpe [21] and by Ngai et al. [6], who also performed friction tests. The major finding was that CoF decreased at an early stage of protein deposition, but it was hypothesized to increase in time because of protein denaturation. Protein deposition on contact lenses was also found to be related to comfort, vision and increased inflammatory responses, as reviewed by Jones [22]. In this study, the physiology of the eye was taken into account by establishing a measuring protocol that mimics the natural environment. For this optimization step, not only parameters such as ageing and sliding speed were considered but also the role of eyelid and tears. The eyelid surface was mimicked by the counter surface, which was required to meet several conditions. It had to be hard and flat so that any lens deformation would be accommodated on contact. This would yield a constant contact area and eliminate the modeling required when using a spherical counter surface [4]. For this study, a hard–soft contact pair was chosen rather than the soft–soft contact pair found in the ocular environment. The hard–soft pair was chosen because the goal of this study was to provide a method to conveniently compare materials rather than to mimic the eye’s mechanics exactly. It was felt that the hard glass surface would produce more reproducible results and has less experimental difficulty than a soft counter surface. Additionally, it was required not to adhere components from the contact lens and to provide a surface representative on the ocular environment. The various glass functionalization procedures that were investigated were deliberately chosen with the aim of mimicking the eyelid. PLL-g-PEG, well known for its lubricating properties in an aqueous environment [23], gave rise to the lowest CoF,

123

394

Tribol Lett (2011) 44:387–397

Table 4 Coefficients of friction measured on daily disposable contact lenses at 0.1 mm/s, arranged in ascending order of CoF at 100 cycles Name of contact lens TM

CLARITI

1 day 

1•DAY ACUVUE MOIST

TM

0 Cycles

Standard deviation (±)

50 Cycles

Standard deviation (±)

100 Cycles

Standard deviation (±)

0.017

0.001

0.014

0.002

0.017

0.004

0.022

0.010

0.019

0.006

0.024

0.003

(narafilcon B)

0.032

0.013

0.030

0.006

0.034

0.009

(narafilcon A)

0.031

0.028

0.035

0.021

0.037

0.019

0.046

0.023

0.024

0.004

0.047

0.029

0.122

0.021

0.113

0.016

0.091

0.009

Proclear 1 day TM ClearSight 1 Day

0.090 0.078

0.033 0.018

0.081 0.141

0.039 0.036

0.125 0.181

0.073 0.037

DAILIES AquaComfort Plus

0.344

0.000

0.474

0.012

0.424

0.051

1•DAY ACUVUE TruEye

TM

1•DAY ACUVUE TruEye

TM

1•DAY ACUVUE

 TM

Focus DAILIES All day comfort 

Table 5 Coefficients of friction measured on reusable contact lenses at 0.1 mm/s, arranged in ascending order of CoF at 100 cycles Name of contact lens

0 Cycles

Standard deviation (±)

50 Cycles

Standard deviation (±)

100 Cycles

Standard deviation (±)

AVAIRA

0.011

0.002

0.018

0.003

0.018

0.001

0.016

0.014

0.024

0.018

0.018

0.015

0.034

0.014

0.023

0.005

0.022

0.009

0.022

0.005

0.021

0.001

0.024

0.004

0.029

0.002

0.028

0.002

0.042

0.010



ACUVUE OASYS

TM

TM

CLARITI

ACUVUE ADVANCE

TM

ACUVUE ADVANCE

TM

PLUS

Biofinity

0.033

0.009

0.057

0.004

0.050

0.002



0.093 0.108

0.023 0.051

0.069 0.110

0.010 0.093

0.090 0.166

0.010 0.058

PremiO

0.177

0.010

0.207

0.023

0.176

0.012

AIR OPTIX AQUA

0.178

0.061

0.187

0.045

0.222

0.073

AIR OPTIX (formerly O2 OPTIX )

0.343

0.038

0.361

0.048

0.292

0.031

NIGHT & DAY

0.391

0.081

0.394

0.040

0.382

0.027



0.415

0.037

0.443

0.041

0.423

0.032

38

0.203

0.045

0.587

0.010

0.551

0.002

SOFLENS

0.562

0.063

0.542

0.033

0.513

0.017

ACUVUE [AIR OPTIX] NIGHT & DAY AQUA

TM

PureVision TM

Optima

whereas hydrophilic glass led to an unusual behavior (Fig. 8). Against a hydrophilic counter surface, 1•DAY TM ACUVUE MOIST brand contact lenses exhibited a higher friction coefficient than 1•DAY ACUVUE despite the presence of polyvinylpyrrolidone (PVP). This is inconsistent with the tactile sensation felt when rubbing TM 1•DAY ACUVUE MOIST lens surface between the  fingers. 1•DAY ACUVUE brand does not feel as slippery. Belyakova et al. [24] reported a strong attraction between PVP and silicon dioxide. The higher coefficient of TM friction observed for 1•DAY ACUVUE MOIST brand lens is attributed to an attractive interaction between PVP TM in the 1•DAY ACUVUE MOIST brand lens matrix and hydrophilic glass surface. Therefore, the high tangential force results from adsorption of the PVP on the counter surface rather than from an intrinsically high coefficient of TM friction for 1•DAY ACUVUE MOIST brand. In the case of PLL-g-PEG, no difference in terms of friction could

123

be observed between 1•DAY ACUVUE and 1•DAY TM ACUVUE MOIST brand, most likely masked by the excellent lubricating properties of PEG. For these reasons, the first two functionalization protocols were discarded. All other protocols gave rise to the opposite trend, as expected. Fibrinogen was chosen to have a more biologically oriented counter surface and because it readily adsorbs on glass. The use of lysozyme went one step further, as it is the major component of tears. Both proteins lead to a clearly distinguishable behavior of 1•DAY ACUVUE TM and 1•DAY ACUVUE MOIST brand, although friction coefficients were relatively high. Mucin, a natural lubricant additive, became the surface coating of choice for performing friction tests on glass. It was chosen as it is a glycoprotein found in the mucous membrane that constitutes the palpebral conjunctiva, and is therefore in contact with lenses in their operating environment. In particular, mucin from bovine submaxillary glands was used as it is a

Tribol Lett (2011) 44:387–397

395

Fig. 9 Comparative study: CoF for various commercially available daily disposable contact lenses tested using the optimized protocol, which is tear-mimicking solution as lubricant and mucin-coated silanized glass as counter surface. Contact lenses are represented in ascending (mean of 0, 50 and 100 cycles) friction coefficient order, according to Table 4. Error bars are standard deviation of three repeat experiments

Fig. 10 Comparative study: CoF for various commercially available reusable contact lenses tested using the optimized protocol, which is tearmimicking solution as lubricant and mucin-coated silanized glass as counter surface. Contact lenses are represented in ascending friction coefficient order, according to Table 5. Error bars are standard deviation of three repeat experiments

readily available mucin that functions at pH 7, as do mucins in the eye. Hydrophobization of glass disks before incubation in mucin solution was performed to enhance mucin adsorption. The major finding with these experiments is that the CoF can significantly vary according to the measuring protocol. Another important aspect of the physiology of the eye is the presence of tears, which play a fundamental role in lubrication. For the evaluation of different contact lens materials, the use of a lubrication solution that is closely related to human tears is preferred. Human tears contain a protein array that is similar to serum but is lacking lactoferrin, lysozyme and tear-specific prealbumin [25]. Yoon et al. [26] reported the use of diluted antilogous serum as

eye drops for severely diseased patients. In their studies, they diluted the serum 1:20 with saline, and did not add any of the tear-specific proteins. Hill et al. [27] described a synthetic tear fluid of 5% blood plasma supplemented with lysozyme at 4.5 g/L and lactoferrin at 1.7 g/L. The tearmimicking fluid used in this study was modified increasing the lysozyme to 5 g/L and eliminating the lactoferrin. Lipids are an important part of the dacruon [28] and have been shown to adsorb onto contact lenses [29]. Cher [28] has proposed dacruon as a new description of the tear film. The historical model describes the tear film as a lipid layer, an aqueous layer and a mucus layer. The dacruon model describes it as a continuous concentration gradient of mucin from the ocular surface to the lipid layer of the

123

396

tear film. Lipids are reported to be floating on the surface of the dacruon [28], and therefore would not be involved in the lubrication of the lid on the contact lens unless a break in the tear film would occur, allowing lipids to adsorb on the surface of the lens. This eventuality can take place as a consequence of eye dryness, irritation and on insertion of the contact lens. Our model does not contemplate such situations, but rather focuses on mimicking the eyelid– contact lens system in standard conditions, where the tear film is intact and lipids are not adsorbing on the contact lens surface. Moreover, the inclusion of lipids into the tearmimicking fluid results in a heterogeneous lubricating fluid, which could affect the reproducibility of measurements. For these reasons, lipids were not included in the described model. For the determination of our eye-mimicking system, contact pressure and sliding speed were also taken into account. Typical in vivo pressure and speed values are reported to be around 3–5 kPa and 1.2 mm/s, respectively. We chose normal loads that allow our setup to be in the contact pressure range described in literature. Contact pressures for 1•DAY ACUVUE were estimated by using the Hertzian contact model and were found to be in the range of 2.6–6.5 kPa for the normal loads used in this study. The purpose of this model is to evaluate the coefficient of friction of contact lenses against a realistic counter surface. The data in Fig. 7 suggest that there is a convergence of the coefficients of friction as the sliding speed is increased. This is thought to be the result of an increased importance of hydrodynamic lubrication. However, the goal is to determine the materials properties, and this can be achieved by measuring the coefficient of friction in the boundary–lubrication regime. The sliding speed of 0.1 mm/s, which is one order of magnitude lower than literature values, was chosen to reduce the possibility of hydrodynamic effects and to ensure that measurements were made in the boundary–lubrication region. Once optimization of the measurement protocol was achieved, with respect to the physiological environment, commercially available contact lenses were tested for frictional properties and compared with each other (Figs. 9 and 10). PVP-containing lenses exhibited the lowest friction coefficients, typically less than 0.050, whereas friction coefficient values for non-PVP-containing lenses ranged from TM 0.100 to 0.600. The exceptions are Sauflon CLARITI , which contains the monomer N-vinylpyrrolidone (NVP), and AVAIRA, which contains a monomer analogous TM to NVP. However, Sauflon (CLARITI ) claims that the N-vinylpyrrolidone in the formulation homopolymerizes to form PVP in situ [30]. In the United States Adopted Names Council (USAN) document for AVAIRA (enfilcon A), it is reported that the lens contains a homopolymer of N-etheny-N-methylacetamide, which can homopolymerize

123

Tribol Lett (2011) 44:387–397

to produce a PVP analog in situ. This frictional behavior was expected because of the lubricating properties of PVP or its analogs and confirmed the validity of our measurement protocol, which proved to be successful in distinguishing between contact lens frictional properties.

5 Conclusion Friction tests were run on commercially available contact lenses with a microtribometer. The measurement protocol was optimized according to sliding speed, counter surface and lubricant–solution composition, to achieve a set-up that mimics the physiological environment. The best combination was found to consist of 0.1 mm/s sliding speed, mucin-coated glass as a counter surface and a lubricant based on packing solution containing lysozyme and serum. Measuring contact lens frictional properties without taking physiological conditions into account was shown to lead to different and potentially misleading results, demonstrating the importance of establishing a broadly valid biologically relevant measurement protocol. Acknowledgments This study was supported by VISTAKON, a division of Johnson & Johnson Vision Care Inc., Jacksonville, FL, USA.

References 1. Nichols, J.J.: Contact lenses 2009. Cont. Lens Spectr. 1, 24–32 (2010) 2. Rumpakis, J.M.B.: New data on contact lens dropouts: an international perspective. Rev. Optom. 147(1), 37–42 (2010) 3. Young, G.: Exploring the relationship between materials and ocular health and comfort. Cont. Lens Spectr. 22, 37–40 (2007) 4. Rennie, A.C., Dickrell, P.L., Sawyer, W.G.: Friction coefficient of soft contact lenses: measurements and modeling. Tribol. Lett. 18(4), 499–504 (2005) 5. Nairn, J.A., Jiang, T.: Measurement of the friction and lubricity properties of contact lenses. Proceedings of ANTEC’95, Boston (1995) 6. Ngai, V., Medley, J.B., Jones, L., Forrest, J., Teichroeb, J.: Friction of contact lenses: silicone hydrogel versus conventional hydrogel. Tribol. Interface Eng. Ser 48, 371–379 (2005) 7. Dunn, A.C., Cobb, J.A., Kantzios, A.N., Lee, S.J., Sarntinoranont, M., Tran-Son-Tay, R., Sawyer, W.G.: Friction coefficient measurement of hydrogel materials on living epithelial cells. Tribol. Lett. 30(1), 13–19 (2008) 8. Van Beek, M., Jones, L., Sheardown, H.: Hyaluronic acid containing hydrogels for the reduction of protein adsorption. Biomaterials 29(7), 780–789 (2008) 9. Young, G., Keir, N., Hunt, C., Woods, C.A.: Clinical evaluation of long-term users of two contact lens care preservative systems. Eye Cont. Lens Sci. Clin. Pract. 35(2), 50–58 (2009) 10. Simmons, P.A., Donshik, P.C., Kelly, W.F., Vehige, J.G.: Conditioning of hydrogel lenses by a multipurpose solution containing an ocular lubricant. Cont. Lens Assoc. Ophthalmol. J. 27(4), 192–194 (2001) 11. Barabino, S., Rolando, M., Camicione, P., Chen, W., Calabria, G.: Effects of a 0.9% sodium chloride ophthalmic solution on the

Tribol Lett (2011) 44:387–397

12.

13.

14.

15.

16.

17.

18.

19.

20. 21.

ocular surface of symptomatic contact lens wearers. Can. J. Ophthalmol. 40(1), 45–50 (2005) Santodomingo-Rubido, J., Barrado-Navascues, E., Rubido-Crespo, M.J.: Ocular surface comfort during the day assessed by instant reporting in different types of contact and non-contact lens wearers. Eye Cont. Lens Sci. Clin. Pract. 36(2), 96–100 (2010) Ozkan, J.J., Snoxall, B., Maher, A., Papas, E.: Lubricants and their effect on comfort with silicone hydrogel and conventional hydrogel lens wear. Invest. Ophthalmol. Vis. Sci. 45, 3161–3164 (2004) Fornasiero, F., Prausnitz, J.M., Radke, C.J.: Post-lens tear-film depletion due to evaporative dehydration of a soft contact lens. J. Membr. Sci. 275(1–2), 229–243 (2006) Shaw, A.J., Collins, M.J., Davis, B.A., Carney, L.G.: Eyelid pressure and contact with the ocular surface. Invest. Ophthalmol. Vis. Sci. 51(4), 1911–1917 (2010) Conway, H.D., Richman, M.: Effects of contact lens deformation on tear film pressures induced during blinking. Am. J. Optom. Physiol. Opt. 59(1), 13–20 (1982) Dong, J., Haugstad, G.D.: Tribology study of PVA contact lens in ionic aqueous environments. Proceedings of American Chemical Society, National Meeting, Polymer Preprints, Baltimore (2005) Lydon, F., Benning, B., Young, R., Tighe, B.J.: Frictional behaviour of contact lenses and ophthalmic solutions: measurement and clinical consequences. Contact Lens Ant. Eye. 25, 35 (2002) Chaudhri, M.M., Yoffe, E.H.: The area of contact between a small sphere and a flat surface. Philos. Mag. A. 44(3), 667–675 (1981) French, K.: Contact lens material properties part 2—mechanical behavior and modulus. Optician. 230(6026), 29–34 (2005) Garrett, Q., Milthorpe, B.K.: Human serum albumin adsorption on hydrogel contact lenses in vitro. Invest. Ophthalmol. Vis. Sci. 37(13), 2594–2602 (1996)

397 22. Jones, L.: Modern contact lens materials: a clinical performance update. Cont. Lens Spectr. 17, 24–35 (2002) 23. Lee, S., Mu¨ller, M., Ratoi-Salagean, M., Vo¨ro¨s, J., Pasche, S., De Paul, S.M., Spikes, H.A., Textor, M., Spencer, N.D.: Boundary lubrication of oxide surfaces by poly(L-lysine)-g-poly(ethylene glycol) (PLL-g-PEG) in aqueous media. Tribol. Lett. 15(3), 231–239 (2003) 24. Belyakova, L., Varvarin, A., Lyashenko, D., Roik, N.: Study of interaction of poly(1-vinyl-2-pyrrolidone) with a surface of highly dispersed amorphous silica. J. Coll. Interface Sci. 264(1), 2–6 (2003) 25. Janssen, P.T., Bijsterveld, O.P.V.: Origin and biosynthesis of human tear fluid proteins. Invest. Ophthalmol. Vis. Sci. 24(5), 623–630 (1983) 26. Yoon, K.C., Heo, H., Im, S.K., You, I.C., Kim, Y.H., Park, Y.G.: Comparison of autologous serum and umbilical cord serum eye drops for dry eye syndrome. Am. J. Optom 44(1), 86–92 (2007) 27. Hill, G.A., Molock, F.F., Jozefowicz, M., Rathore, O., Jozefonvicz, J., Fadli, Z.: Antimicrobial coatings for ophthalmic devices. US Patent 20040208983A1, 21 October 2004 28. Cher, I.: A new look at lubrication of the ocular surface: fluid mechanics behind the blinking eyelids. Ocul. Surf. 6(2), 79–86 (2008) 29. Carney, F.P., Nash, W.L., Sentell, K.B.: The adsorption of major tear film lipids in vitro to various silicone hydrogels over time. Invest Ophthalmol Vis Sci. 49(1), 120–124 (2008) 30. Broad, R.A.: Contact lens are formed of a composition comprising reaction product of silicone-containing monomer; 3-methacryloxypropyl tris(trimethylsiloxy)silane; N-vinyl pyrrolidone; and other non-ionic hydrophilic monomer. World Intellectual Property Organization WO2008061992-A2, 29 May 2008

123