Magnetic particle imaging for aerosol-based magnetic targeting

This content has been downloaded from IOPscience. Please scroll down to see the full text. 2017 Jpn. J. Appl. Phys. 56 088001 (http://iopscience.iop.org/1347-4065/56/8/088001) View the table of contents for this issue, or go to the journal homepage for more Download details: IP Address: 107.173.143.31 This content was downloaded on 01/07/2017 at 14:10 Please note that terms and conditions apply.

You may also be interested in: Development of a system for magnetic particle imaging using neodymium magnets and gradiometer Kenya Murase, Samu Hiratsuka, Ruixiao Song et al. Simple and practical method for correcting the inhomogeneous sensitivity of a receiving coil in magnetic particle imaging Kenya Murase, Natsuo Banura, Atsushi Mimura et al. Experiments with neodymium magnets David Featonby Magnetic particle imaging: from proof of principle to preclinical applications T Knopp, N Gdaniec and M Möddel Combining magnetic particle imaging and magnetic fluid hyperthermia in a theranostic platform Daniel Hensley, Zhi Wei Tay, Rohan Dhavalikar et al. Magnetic cell targeting with MRI J Riegler, B Allain, R J Cook et al. Characterization of patient inhaler inhalation sounds using non-contact and tracheal microphones Terence E Taylor, Céline De Looze, Philippa MacHale et al. In vitro and in vivo comparison of a tailored magnetic particle imaging blood pool tracer with Resovist Michael Gerhard Kaul, Tobias Mummert, Caroline Jung et al. Retention assessment of magnetic nanoparticles in rat arteries with micro-computed tomography Shu-Ju Tu, Siao-Yun Wu, Fu-Sheng Wang et al.

BRIEF NOTE

Japanese Journal of Applied Physics 56, 088001 (2017) https://doi.org/10.7567/JJAP.56.088001

Magnetic particle imaging for aerosol-based magnetic targeting Natsuo Banura and Kenya Murase* Department of Medical Physics and Engineering, Division of Medical Technology and Science, Faculty of Health Science, Graduate School of Medicine, Osaka University, Suita, Osaka 565-0871, Japan *E-mail: [email protected] Received April 29, 2017; accepted May 29, 2017; published online June 30, 2017 Magnetic targeting is a strategy for improving the efficacy of therapeutic agents and minimizing the unwanted side effects by attaching the therapeutic agents to magnetic nanoparticles (MNPs) and concentrating them to the targeted region such as solid tumors and regions of infection using external magnetic fields. This study was undertaken to investigate the usefulness of magnetic particle imaging (MPI) for monitoring the effect of aerosol-based magnetic targeting by phantom experiments using a simple flow model and nebulized MNPs. Our results suggest that MPI is useful for monitoring the effect of aerosol-based magnetic targeting. © 2017 The Japan Society of Applied Physics

Magnetic nanoparticles (MNPs) have been used in various medical imaging fields such as magnetic resonance imaging (MRI) and magnetic particle imaging (MPI). MPI is an imaging method that was introduced in 2005.1) MPI allows the imaging of the distribution of MNPs with high sensitivity, high spatial resolution, and high imaging speed.1) In addition, MPI can visualize MNPs in a positive contrast with no signals from background tissues and can quantify the number of MNPs with excellent linearity.2) These advantages have been shown to be useful for various medical applications such as vascular imaging,3) stem cell tracking,4) cancer imaging,5) and magnetic hyperthermia.6) Magnetic targeting has been developed for localizing drug carriers such as liposomes containing both MNPs and therapeutic agents in the targeted organ or tissue by applying external magnetic fields.7) The magnetic targeting of therapeutic agents results in the concentration of the agents at the targeted site, consequently reducing or eliminating systemic unwanted side effects.7) Recently, a new magnetic targeting strategy called “aerosol-based magnetic targeting” has been introduced for lung diseases.8) Aerosol-based magnetic targeting uses magnetic aerosol droplets comprising nebulized MNPs and therapeutic agents.8) The targeted delivery of magnetic aerosol droplets to the lung has various inherent advantages in that the drug dose can be easily adjusted by changing the drug concentration in the aerosol droplets without binding the drug to MNPs and that several drugs can be applied simultaneously using only one MNP type. The most important feature in the aerosol-based drug delivery to the lung is the direct application of the drug in the liquid phase to the disease lesion in the lung, which leads to a fast onset of pharmaceutical action and reduced systemic side effects.8) An accurate knowledge of the spatial distribution and number of magnetic aerosol droplets is crucial for enhancing the effect of aerosol-based magnetic targeting and for its optimization. We previously investigated the application of MPI to pulmonary imaging using nebulized MNPs and attempted to quantify the mucociliary clearance in the lung using MPI.9) To the best of our knowledge, however, the usefulness of MPI for monitoring the effect of aerosol-based magnetic targeting has not been investigated yet. In this study, we investigated it using a simple flow phantom simulating aerosol-based magnetic targeting.

Fig. 1. Schematic illustration of a flow phantom simulating aerosol-based magnetic targeting. d represents the distance from the magnet surface to the silicon tube surface.

Figure 1 illustrates our flow phantom simulating aerosolbased magnetic targeting. As shown in Fig. 1, our phantom consists of a straight silicon tube (16 mm in outer diameter, 12 mm in inner diameter, and 50 mm in length), a compressor-type nebulizer with a suspension reservoir (Omron NE-C28), a flow meter (Kofloc RK1710), an inhaler (Shin-Ei Industries Minic S-II), and a neodymium magnet (Sangyo Supply). The MNPs in the reservoir were nebulized using the compressor-type nebulizer and pumped out from the inlet of the silicon tube to the outlet using the inhaler. In this study, we used 5-fold-diluted M300 (Fe3O4, iron concentration: 1.2 mol=L and particle size: 17.1 ± 4.9 nm)10) (Sigma HiChemical) as the source of MNPs. A neodymium magnet (diameter: 10 mm, length: 30 mm, and magnetic flux density on the surface: 0.5 T) was placed near the center of the silicon tube to attract the MNPs. Furthermore, we placed a cylindrical urethane sponge (10 mm in width) inside the silicon tube to hold the MNPs accumulating at the targeted site. The urethane sponge also simulates the lung tissue. The flow rate was controlled using the flow meter placed between the outlet of the silicon tube and the inhaler. To image the distribution of MNPs accumulating at the targeted site in the silicon tube using MPI, we took out the silicon tube after the phantom experiments and imaged it using our MPI scanner.11,12) In this study, we used Osaka MPI Scanner II, which is an extended version of our previous scanner.11,12) In this scanner, two opposing neodymium magnets were used for generating a field-free line (FFL). The gradient strengths of the magnetic fields perpendicular and

Content from this work may be used under the terms of the Creative Commons Attribution 4.0 license. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

088001-1

© 2017 The Japan Society of Applied Physics

Jpn. J. Appl. Phys. 56, 088001 (2017)

N. Banura and K. Murase

Fig. 2. (Color online) Images obtained by MPI of the MNPs accumulating in the urethane sponges (Fig. 1) when d was varied at 0, 2, 4, 6, 8, and 10 mm. In these experiments, the nebulized MNPs were inhaled with an inhaler (Fig. 1) for 15 min at a flow rate of 1.5 L=min. The MPI images were superimposed on the X-ray CT images. A neodymium magnet was located at the right side of the image. Scale bar = 5 mm.

parallel to the FFL were 3.9 and 0.1 T=m, respectively. A gradiometer coil was used for receiving the third-harmonic signals generated by MNPs. The frequency and peak-to-peak amplitude of the drive magnetic field were taken as 400 Hz and 20 mT, respectively. After 36 projection data were acquired by translating and rotating a sample in the gradiometer coil automatically, transverse images were reconstructed using the maximum likelihood-expectation maximization (ML-EM) algorithm with 15 iterations.11) For coregistration, X-ray computed tomography (CT) images were also acquired using a four-row multislice CT scanner (Toshiba Medical Systems Asteion) with a tube voltage of 120 kV, a tube current of 210 mA, and a slice thickness of 0.5 mm. After the MPI studies, we drew a region of interest (ROI) on the MPI image and calculated the average MPI value within the ROI by taking the threshold value for extracting the contour of the image as 40% of the maximum MPI value within the ROI. Figure 2 shows the MPI images of the urethane sponges taken out after the phantom experiments, in which the distance from the magnet surface to the silicon tube surface (d ) was varied at 0, 2, 4, 6, 8, and 10 mm. In these experiments, the nebulized MNPs were inhaled with the inhaler for 15 min at a flow rate of 1.5 L=min. The MPI images were superimposed on the X-ray CT images. As shown in Fig. 2, the MPI images were shifted to the magnet side of the silicon tube and the pixel values decreased with increasing d. Figure 3 shows the relationship between d and the average MPI value, in which the nebulized MNPs were inhaled for 15 min at a flow rate of 1.5 L=min. The dotted line shows the average MPI value in the case without a magnet (Fig. 1) [0.064 ± 0.0067 (mean ± standard error for n = 5)]. When regression analysis was performed using a function of y = a · exp(−bd ) + 0.064 with a and b being constants, there was an excellent correlation between them (r = 0.980) and the decay constant (b) was 0.130 mm−1. Figure 4 shows the MPI images of the urethane sponges taken out after the experiments, in which the flow rate was varied at 1.5, 1.0, 0.5, and 0.25 L=min. In these experiments, d was fixed at 4 mm and the duration of inhalation was taken

Fig. 3. Relationship between d and the average MPI value when the nebulized MNPs were inhaled for 15 min at a flow rate of 1.5 L=min. Data are presented as the mean ± standard error (SE) for n = 5. The solid line represents the fitted curve using a function of y = a · exp(−bd ) + 0.064, with a and b being constants. The dotted line shows the average MPI value in the case without a magnet (Fig. 1) (0.064).

Fig. 4. (Color online) MPI images of the MNPs accumulating in the urethane sponges (Fig. 1) when the flow rate was varied at 1.5, 1.0, 0.5, and 0.25 L=min. In these experiments, d was fixed at 4 mm and the nebulized MNPs were inhaled for 15 min. The MPI images were superimposed on the X-ray CT images. A neodymium magnet was located at the right side of the image. Scale bar = 5 mm.

as 15 min. As in Fig. 2, the MPI images were superimposed on the X-ray CT images. As shown in Fig. 4, the pixel values increased with increasing flow rate. Figure 5 shows the relationship between the flow rate and the average MPI value, in which d and the duration of inhalation were taken as 4 mm and 15 min, respectively. When regression analysis was performed using the phenomenological Box–Lucas equation,13) there was an excellent correlation between them (r = 0.988). As shown in Fig. 5, the average MPI value increased but tended to saturate with increasing flow rate. In this study, we investigated the usefulness of MPI for monitoring the effect of aerosol-based magnetic targeting using a simple flow phantom (Fig. 1) and nebulized MNPs. As shown in Figs. 2–5, the changes in the MPI image and average MPI value depending on d and flow rate were clearly observed. When we investigated the correlation between the iron concentration of MNPs (x in mg=mL) and the average MPI value (y) using phantoms, there was an excellent linear correlation between them with a correlation coefficient greater than 0.999 and a regression equation of y = 0.183x + 0.0535 (plot not shown), indicating that the average

088001-2

© 2017 The Japan Society of Applied Physics

Jpn. J. Appl. Phys. 56, 088001 (2017)

N. Banura and K. Murase

Fig. 5. Relationship between the flow rate and the average MPI value when d was fixed at 4 mm and the nebulized MNPs were inhaled for 15 min. Data are presented as the mean ± SE for n = 5. The solid line represents the fitted curve using the phenomenological Box–Lucas equation.13)

MPI value can be converted to the iron concentration in mg=mL using this regression equation. Combining this finding and the present results (Figs. 2–5) suggests that MPI allows us to visually and quantitatively evaluate the effect of aerosol-based magnetic targeting. It is well known that the magnetic force acting on MNPs is proportional to the gradient strength of the magnetic field,14) which decays exponentially with increasing d.7) This appears to be the main reason why the average MPI value decreased exponentially with increasing d (Fig. 3). When we measured the gradient strength of the magnetic field perpendicular to the face of the magnet (Fig. 1) along the central axis using a gauss meter and fitted it using an exponential function, the decay constant was 0.162 mm−1. This value was similar to that (0.130 min−1) shown in Fig. 3. As shown in Fig. 5, the average MPI value increased but tended to saturate with increasing flow rate. The increase in average MPI value appears to be due to the fact that the number of MNPs passing through the cross section of the silicon tube is proportional to the flow rate. The tendency for the average MPI value to saturate with increasing flow rate would be explained by the relationship between the mean velocity of MNPs and the magnetic force acting on then. When there exists the magnetic force acting on an MNP with the magnetic moment (m), the so-called “magnetic” velocity perpendicular to the flow direction (vm) is given by14) vm ¼

m @B ; 3dh @r

ð1Þ

where η is the viscosity of the medium, dh is the hydrodynamic diameter of the MNP, and ∂B=∂r is the magnetic field gradient perpendicular to the flow direction. At a mean flow velocity (v), the time (tF) required for an MNP to cross a magnetic field with a usable field range (LF) is given by14) tF ¼

LF : v

ð2Þ

Thus, the moving distance of the MNP during tF perpendicular to the flow direction (Δr) is given by r ¼ vm  tF ¼

mLF @B ; 3dh v @r

ð3Þ

indicating that Δr is inversely proportional to v . Although the number of MNPs passing through the cross section of the silicon tube is proportional to the flow rate, the moving distance of the MNP to the magnet side (Δr) is inversely proportional to v or flow rate as given by Eq. (3). This appears to be the main reason why there was a tendency for the average MPI value to saturate with increasing flow rate (Fig. 5). As previously described, when investigating the relationship between d and the average MPI value (Figs. 2 and 3), the nebulized MNPs were inhaled for 15 min at a flow rate of 1.5 L=min, implying that the total inhaled volume was 22.5 L. We adopted the above values for inhalation duration and flow rate in consideration of both the capacity of our inhaler and the sensitivity of our MPI scanner. It has been reported that the tidal volume per respiration in healthy adults is approximately 0.5 L.15) When the inspiratory duration is 1 s, the inspiratory flow rate becomes 30 L=min, which is much greater than our flow rate. When the respiratory frequency is assumed to be 15 min−1, the above total inhaled volume (22.5 L) corresponds to the tidal volume summed up for 3 min or 45 respirations. For the investigation of more realistic cases, it would be necessary to increase the flow rate and shorten the inhalation duration. A limitation of this study is that the MPI value was obtained from a single slice of the MPI image. Thus, the analysis using a single slice of the MPI image limits the accurate evaluation of the spatial distribution of MNPs and the amount of iron in the entire targeted region. For a more detailed analysis, it will be necessary to acquire three-dimensional data, evaluate the three-dimensional distribution of MNPs, and quantify the amount of iron in the targeted region from these data. In summary, the present study suggests that MPI is useful for monitoring the effect of aerosol-based magnetic targeting. Acknowledgements This work was supported by Grants-in-Aid for Scientific Research (Grant Nos. 25282131 and 15K12508) from the Japan Society for the Promotion of Science and Japan Science and Technology Agency.

1) B. Gleich and J. Weizenecker, Nature 435, 1214 (2005). 2) K. Murase, M. Aoki, N. Banura, K. Nishimoto, A. Mimura, T. Kuboyabu, and I. Yabata, Open J. Med. Imaging 5, 85 (2015). 3) J. Weizenecker, B. Gleich, J. Rahmer, H. Dahnke, and J. Borgert, Phys. Med. Biol. 54, L1 (2009). 4) J. W. M. Bulte, P. Walczak, M. Janowski, K. M. Krishnan, H. Arami, A. Halkola, B. Gleich, and J. Rahmer, Tomography 1, 91 (2015). 5) E. Y. Yu, M. Bishop, B. Zheng, R. M. Ferguson, A. P. Khandhar, S. J. Kemp, K. M. Krishnan, P. W. Goodwill, and S. M. Conolly, Nano Lett. 17, 1648 (2017). 6) K. Murase, J. Oonoki, H. Takata, R. Song, A. Angraini, P. Ausanai, and T. Matsushita, Radiol. Phys. Technol. 4, 194 (2011). 7) T. Kuboyabu, A. Ohki, N. Banura, and K. Murase, Open J. Med. Imaging 6, 33 (2016). 8) P. Dames, B. Gleich, A. Flemmer, K. Hajek, N. Seidl, F. Wiekhorst, D. Eberbeck, I. Bittmann, C. Bergemann, T. Weyh, L. Trahms, J. Rosenecker, and C. Rudolph, Nat. Nanotechnol. 2, 495 (2007). 9) K. Nishimoto, A. Mimura, M. Aoki, N. Banura, and K. Murase, Open J. Med. Imaging 5, 49 (2015). 10) K. Murase, T. Konishi, Y. Takeuchi, H. Takata, and S. Saito, Radiol. Phys. Technol. 6, 399 (2013). 11) K. Murase, S. Hiratsuka, R. Song, and Y. Takeuchi, Jpn. J. Appl. Phys. 53, 067001 (2014). 12) K. Murase, R. Song, and S. Hiratsuka, Appl. Phys. Lett. 104, 252409 (2014). 13) G. E. P. Box and H. L. Lucas, Biometrika 46, 77 (1959). 14) B. Gleich, N. Hellwig, H. Bridell, R. Jurgons, C. Seliger, C. Alexiou, B. Wolf, and T. Weyh, IEEE Trans. Nanotechnol. 6, 164 (2007). 15) E. P. Radford, J. Appl. Physiol. 7, 451 (1955).