Relative Contribution of Wall Shear Stress and ... - Semantic Scholar

5 downloads 0 Views 244KB Size Report
The Biomedical Engineering and. Mathematics Programs,. Louisiana Tech University,. Ruston, LA. Seymour Glagov. The Departments of Surgery and Pathology ...
Francis Loth Departments of Mechanical Engineering and Bioengineering, University of Illinois at Chicago, Chicago, IL

Steven A. Jones The Biomedical Engineering Program, Louisiana Tech University, Ruston, LA

Christopher K. Zarins The Parker H. Petit Institute for Bioengineering and Bioscience, Georgia Tech/Emory, Atlanta, GA

Don P. Giddens The Department of Surgery, Stanford University, Stanford, CA

Raja F. Nassar The Biomedical Engineering and Mathematics Programs, Louisiana Tech University, Ruston, LA

Seymour Glagov The Departments of Surgery and Pathology, The University of Chicago, Chicago, IL

Hisham S. Bassiouny The Department of Surgery, The University of Chicago, Chicago, IL

Relative Contribution of Wall Shear Stress and Injury in Experimental Intimal Thickening at PTFE End-to-Side Arterial Anastomoses Background: Intimal hyperplastic thickening (IHT) is a frequent cause of prosthetic bypass graft failure. Induction and progression of IHT is thought to involve a number of mechanisms related to variation in the flow field, injury and the prosthetic nature of the conduit. This study was designed to examine the relative contribution of wall shear stress and injury to the induction of IHT at defined regions of experimental end-to-side prosthetic anastomoses. Methods and Results: The distribution of IHT was determined at the distal end-to-side anastomosis of seven canine Iliofemoral PTFE grafts after 12 weeks of implantation. An upscaled transparent model was constructed using the in vivo anastomotic geometry, and wall shear stress was determined at 24 axial locations from laser Doppler anemometry measurements of the near wall velocity under conditions of pulsatile flow similar to that present in vivo. The distribution of IHT at the end-to-side PTFE graft was determined using computer assisted morphometry. IHT involving the native artery ranged from 0.0⫾0.1 mm to 0.05⫾0.03 mm. A greater amount of IHT was found on the graft hood (PTFE) and ranged from 0.09⫾0.06 to 0.24⫾0.06 mm. Nonlinear multivariable logistic analysis was used to model IHT as a function of the reciprocal of wall shear stress, distance from the suture line, and vascular conduit type (i.e. PTFE versus host artery). Vascular conduit type and distance from the suture line independently contributed to IHT. An inverse correlation between wall shear stress and IHT was found only for those regions located on the juxta-anastomotic PTFE graft. Conclusions: The data are consistent with a model of intimal thickening in which the intimal hyperplastic pannus migrating from the suture line was enhanced by reduced levels of wall shear stress at the PTFE graft/host artery interface. Such hemodynamic modulation of injury induced IHT was absent at the neighboring artery wall. 关DOI: 10.1115/1.1428554兴 Keywords: hemodynamics, anastomosis, intimal hyperplasia, wall shear stress, fluid dynamics, end-to-side vascular graft

Introduction Progressive intimal hyperplastic thickening 共IHT兲 remains a major cause of intermediate and long term failure of bypass grafts 关1–5兴 and endovascular interventions 关6 – 8兴. Anastomotic IHT has been associated with hemodynamic factors 关9兴, compliance mismatch between graft material and host vessel 关10兴, biomaterials incompatibility 关11兴, and mural injury 关12兴. However, quantitative relationships with any given factor have not been reported. Previous studies have demonstrated wall shear stress plays a critical role in adaptive wall remodeling. Such adaptive responses tend to normalize wall shear stress to levels between 10 and 20 dynes/cm2 关13–15兴. A similar increase in intimal thickness along polytetrafluoroethylene 共PTFE兲 grafts in response to reduced hemodynamic wall shear stress has also been observed 关16兴. However, the role of hemodynamic wall shear stress in modulating IHT in a relatively confined region such as the end-to-side anastomosis is yet unclear. The hemodynamic wall shear stress in this region is characterized by dramatic spatial variation of shear stress in both magnitude and direction. A number of studies have suggested that wall shear stress or Contributed by the Bioengineering Division for publication in the JOURNAL OF BIOMECHANICAL ENGINEERING. Manuscript received by the Bioengineering Division May 26, 2000; revised manuscript received September, 17 2001. Associate Editor: C. Ross Ethier.

44 Õ Vol. 124, FEBRUARY 2002

other hemodynamic forces may enhance the development of anastomotic IHT. Imparato 关1兴 suggested that the locations of obstructive lesions in venous bypass grafts and atherosclerotic plaques in arteries may have a ‘‘common denominator of specific hemodynamic conditions.’’ Sottiurai et al. 关17兴 reported that IHT in endto-side PTFE iliofemoral grafts in dogs was located at the heel and toe of the graft and the floor of the host artery. Based on these observations and independent measurements of wall shear stress in an in vitro model of an end-to-side anastomosis, Ojha suggested that low wall shear stress 关18兴 or high wall shear stress gradients 关19兴 might be responsible for the localization of anastomotic IHT. Numerical correlation between these in vivo and in vitro results could not be obtained because the in vivo data were only qualitative. Furthermore, standardization of the in vivo geometry is necessary to adequately compare the in vivo biological and the in vitro fluid dynamic data since geometric and flow variations strongly affect the hemodynamics environment 关20兴. Recently Keynton et al. 关21兴 have studied a model in which both IHT and wall shear stress were measured in vivo. They found a small but significant correlation between IHT and hemodynamic variables. We have previously shown that regions of IHT in implanted canine iliofemoral PTFE and autologous vein grafts corresponded to locations of slow particle velocities in a scaled up in vitro model 关22,23兴. Increasing the flow rate over two-fold in experimental canine anastomoses by creation of a distal arterial-venous

Copyright © 2002 by ASME

Transactions of the ASME

fistula virtually eliminated anastomotic IHT at the suture line 关23兴. This is consistent with increased near-wall particle velocities in the in vitro flow model 关24兴. Furthermore, outflow limitation in a model of anastomotic intimal hyperplasia in a New Zealand White adult male rabbit enhanced progression of IHT to an occlusive lesion 关25兴. The findings of these studies suggest that anastomotic IHT is likely to occur in regions of low and oscillatory flow, but the qualitative nature of the in vitro data has not yet allowed quantitative comparisons of wall shear stress with IHT. In order to establish the quantitative relationship between IHT at distal anastomoses and its hemodynamic environment, we conducted a carefully controlled study to measure IHT inside a canine model in parallel with fluid dynamics simulations of the same geometry. The correlation coefficients for this relationship were much lower than anticipated from previous qualitative observations. We propose that the difficulty in establishing such a relationship is caused by the interplay between multiple biological and mechanical factors 共i.e. hemodynamics, surgical injury, and the presence of PTFE兲. In this paper, we present the results of our study and examine the statistical relationships between IHT, reciprocal of wall shear stress, distance from the suture line, and the nature of the vascular conduit 共PTFE versus native artery wall兲. The comparison is performed by multi-regression with backward elimination 关26兴.

Materials and Methods I

Acute in Vivo Studies

Animal Model. Acute in vivo experiments were conducted to obtain the baseline blood flow rates and the graft geometry for the in vitro flow study. Adult male mongrel dogs (n⫽4) weighing 20 to 25 Kg were anesthetized with intravenous sodium pentobarbital 共30 mg/kg兲, and maintained after endotracheal intubation with one percent to three percent halothane. The iliofemoral arterial segments were exposed retroperitoneally through inguinal incisions extending to the upper thigh. Femoral artery diameter was measured with calipers before manipulation. Iliofemoral bypasses were constructed using 6 mm thin-walled PTFE 共30 ␮m internodal distance, reinforced, W. L. Gore Inc., Flagstaff, AZ兲. PTFE was used rather than vein because its diameter was well defined which helped reduce the geometric variability from animal to animal. The nomenclature used is given in Fig. 1. The proximal anastomosis was constructed end-to-end to the common iliac artery, and the distal anastomosis was constructed end-to-side to the femoral artery with a 6-0 Prolene 共Ethicon Inc., Somerville, N.J.兲 suture in a continuous pattern. The external iliac artery was ligated one cm above the inguinal ligament to simulate an arterial occlusion. The distal anastomotic geometry was standardized with a hood-length to vessel diameter ratio between 4:1 and 5:1, based on external artery dimensions, and a maximum hood width 共see Fig. 3兲 of six mm at the center of the anastomosis. To standardize outflow conditions, all distal anastomoses were constructed one cm beyond the deep femoral artery in a segment of superficial femoral artery devoid of side branches. A sketch of the surgical configuration is shown in Fig. 2. Housing and handling of animals

Fig. 1 Nomenclature of the end-to-side anastomosis „POSproximal outlet segment, DOS-distal outlet segment…

Journal of Biomechanical Engineering

Fig. 2 Sketch showing the surgical bypass configuration in which blood bypasses an occlusion „ligation… through the PTFE and exits the distal anastomosis through both the proximal outlet segment „POS… and the distal outlet segment „DOS…

were in compliance with ‘‘Principles of Laboratory Animal Care’’ and ‘‘Guide for the Care and Use of Laboratory Animals’’ 共NIH publication no. 80-23, revised 1985兲. Hemodynamic Measurements. After the grafts were implanted for one hour, blood flow was measured in the common iliac artery proximal to the bypass graft and in the distal outflow segment 共DOS兲 of the distal end-to-side anastomosis with transit time ultrasound flow probes 共Transonics Inc., Ithaca, NY兲. Systemic blood pressure was monitored with an indwelling 21-gauge brachial arterial cannula connected to a Statham pressure transducer and a strip chart recorder 共Gould Inc., Cleveland, Ohio兲. Preliminary studies using four dogs provided the necessary hemodynamic data for the in vitro experiments. Pulsatile blood flow rates were measured by electromagnetic flow meters 共Carolina Medical Electronics兲 and stored on FM tape by a four channel Racal frequency modulated recorder and then analyzed on a Masscomp 5500 microcomputer 共Concurrent, Westford, Mass.兲. II Chronic Animal Studies. Chronic in vivo studies provided the quantitative IHT data. The chronic studies (n⫽9) followed the same procedure as the acute studies. After the operation, all animals received 125 mg of aspirin every day to reduce platelet aggregation and promote graft patency. After 12 weeks, the animals were sacrificed with an intravenous overdose of pentobarbital, and the distal infrarenal aorta was cannulated. Of the nine PTFE grafts, seven were patent at the time of sacrifice 共78 percent patency rate兲. The distal aorta, iliac arteries, bypass grafts, and distal anastomoses were fixed by pressure perfusion with three percent glutaraldehyde at 100 mm Hg. The distal anastomosis, including the graft hood, suture line, and proximal and distal outflow segments 共POS and DOS兲 were harvested en bloc and cross sectioned at 200 micron intervals. Sections were embedded in paraffin, and stained with hematoxylin and eosin, and by the Gomori-trichrome-aldehyde fuchsin procedure for connective tissue differentiation. The resulting slides were examined for the presence of anastomotic IHT by light microscopy. The histological slides were projected on a digitizing plate and IHT was quantitated after tracing the lumen and internal elastic lamina using computer assisted morphometric techniques every millimeter 共i.e., every fifth cross-section兲. For the correlations, the in vitro locations were matched with in vivo locations separated by one sixth of the in vivo hood length for locations within the anastomosis. While variability in hood length was small, the in vivo locations were nondimensionalized by in vivo hood length to insure that all locations were compared to in vitro locations relative to both the toe and heel locations. Otherwise, positions in the POS of the in vivo model would potentially be matched with positions within the anastomosis in the in vitro model. The variability was due to varying amounts of shrinkage during fixation and differences in the hood length and artery diameter from graft to graft. The in vivo measurement positions ranged from 1.8 diameters proximal to the heel to two diameters distal of the toe. In FEBRUARY 2002, Vol. 124 Õ 45

Fig. 3 Regions on the hood and floor over which IHT was averaged circumferentially. Five intimal thickening values, 15 degrees apart, were used in the calculation for IHT for each region.

each histologic section, IHT was averaged circumferentially over ⫾30 degrees centered along the hood or floor as shown in Fig. 3. The mean and standard deviation IHT at each axial position was computed as the average value for the seven dogs. Positions without IHT were entered as zero in the averaging. III

Model Flow Studies

Model Geometry. A laser Doppler anemometer 共TSI system, Lexel four watt Argon ion laser兲 was used to determine the velocity distribution inside an in vitro model of the anastomosis. The probe volume sample size was 0.5 mm in length by 0.05 mm waist diameter 共manufacturer’s specifications兲. The model was constructed from a Sylgard elastomer 共Dow Corning, Midland, MI兲 and scaled up nine times to increase the spatial resolution. Dynamic similarity between the original graft and the scaled-up model was ensured by maintaining physiologic Reynolds and Womersley numbers. The wall shear stress values were estimated from the near wall velocity measurements. The in vitro model incorporated geometric parameters which were measured from a polysiloxane 共3M company, St. Paul, MN兲 cast of an acutely harvested end-to-side anastomosis. These parameters included the graft to host artery lumen diameter ratio 共1.6 to 1兲, the branching angle 共5 degrees兲, the hood length to host artery diameter ratio 共5 to 1, based on internal dimensions兲 and the shapes of seven cross-sections of the cast along the host artery axis 共Fig. 4兲. The cast cross-sections were obtained from a high resolution computerized tomography scanner 共GE 9800兲. The final model was symmetric about the bifurcation plane, had smooth contours at the junctions, and had a curved contour along the hood. Details of the model construction technique have been previously described by Loth et al. 关27,28兴. Axial wall shear stress was measured from 1.2 diameters distal to the toe to a position 6.0 diameters proximal to the toe at intervals of 0.8 diameters. Pulsatile in Vitro Flow System. The model was mounted on a recirculating flow rig with flow entering the graft model via a 46 Õ Vol. 124, FEBRUARY 2002

Fig. 4 Geometry for the in vitro flow model. Wall shear stress measurements were obtained at the axial locations indicated by the dotted lines. The cross-sections of the model at each of these axial locations are indicated.

straight plastic tube three meters in length and 5.08 cm in diameter allowing for fully developed pulsatile flow at the entrance. The fluid was a mixture of 42 percent water and 58 percent glycerin by weight in order to match the refractive index of the Sylgard model. This fluid has a refractive index of 1.41, a density of 1.16 g/ml and a kinematic viscosity of 9.046 centipoise as measured by a Wells-Brookfield cone and plate micro-viscometer at 26.5°C. The flow was driven by a centrifugal pump and modulated by a Target Rock control valve to generate the desired waveform 共Fig. 5兲. The flow rate entering the graft segment and exiting via the proximal outflow segment were monitored by two Carolina Medical Electronics electromagnetic flow meters. The in vitro flow waveform is representative of the femoral artery flow conditions measured during the acute canine experiments after graft stabilization. The mean, positive peak, and reverse peak Reynolds numbers were 222, 850, and 120, respectively, based on the total inflow rate and host artery diameter (ReD⫽4␳Qgraft /D h ␮ ), where D h 共host artery diameter兲 is 0.35 cm. In vivo values of dynamic viscosity, ␮, is 0.035 g/sec-cm, and the fluid density, ␳, is 1.05 g/ml. The frequency of the in vivo waveform was 1.5 Hz 共90 BPM兲 yielding a Womersley number of 3.0 based on the host vessel diameter. The proximal to distal outlet flow division was set at 20:80. Wall Shear Stress Measurements. Near wall velocity measurements were taken at 24 midplane locations along the graft inlet, hood, floor and inside the proximal and distal outflow segments 共Fig. 4兲. Wall shear stress was obtained from the product of the instantaneous velocity gradient at the wall and the fluid coefficient of viscosity. The instantaneous velocity gradient at the wall was estimated from a linear regression of three velocity measurements near the wall. The velocity measurement nearest to the wall ranged from 0.5–2.2 percent of host artery diameter 共0.16 –0.70 mm兲 depending on the quality of the signal. The radial spacing for the other two velocity measurements was 0.8 percent of the host artery diameter 共0.25 mm兲. The accuracy of this wall shear stress technique was determined for a straight tube and the mean values Transactions of the ASME

highest p-value is then omitted from the model and a new fit is performed with the remaining variables. Omission of variables and refitting continue until all coefficients remaining in the model are statistically significant. A total of 24 data points were used in the correlations.

Results I

Fig. 5 Flow waveform used for the in vitro velocity measurements. The flow rate has been scaled to reflect the equivalent in vivo values. POS and DOS are the proximal and distal outflow segments, respectively.

were estimated to be within ⫾10 percent of the theoretical mean value, while the peak values 共both positive and negative兲 were estimated to be within ⫾20 percent 关27,29兴. Values of mean, maximum and minimum wall shear stress were determined. Also computed were the pulse shear index 共PS兲 and oscillatory shear index 共OSI兲 where the PS is defined as the maximum minus the minimum value of wall shear stress and the OSI is defined according to Ku et al. 关30兴. Wall shear stress gradient was not considered since good estimates of this parameter would require more finely spaced measurements.

Chronic Animal Studies

In Vivo Measurements. Hemodynamic and dimensional variables for the chronic in vivo studies were comparable to the scaled values used in the in vitro study. These variables were measured and averaged over the seven anastomoses and are expressed as mean ⫾ standard deviation. The mean blood flow rates in the graft and the distal outflow segment were 173⫾66 ml/min and 136⫾45 ml/min, respectively. This yielded a proximal-to-distal outflow ratio of 19:81. At the time of sacrifice, the mean diameter of the host artery was 3.19⫾0.35 mm based on the average of the POS and DOS diameters of each animal. Thus, the mean graft to host artery ratio was 1.7:1 based on internal dimensions 共internal diameter of the PTFE graft was ⬃5.6 mm兲. The mean hood length of the grafts was 12.9⫾1.9 mm. This gave an average hood length of 4.0⫾0.6 host artery diameters. These measurements show the in vivo graft geometry to be 15–20 percent smaller than the in vitro model geometry after fixation. This difference is caused by shrinkage and is typically found when tissue is fixed with three percent glutaraldehyde 关31兴. The kinematic viscosity was assumed to be 0.035 cm2/sec. The mean Reynolds number was 262 based on the total graft inflow and the host artery diameter. This is 18 percent higher than that of the in vitro studies. Qualitative Topographic and Histopathologic Features of IHT. Both graft hood and suture line IHT were present in all anastomoses. Hood IHT was diffused extending axially from the suture line at the toe to position ⫺6.0, and circumferentially to the suture line along the anastomotic sinus sidewall. IHT also extended from the suture line along the sidewall of the host artery, but did not extend as far as the floor. IHT consisted of smooth muscle cells

IV Statistical Analysis. A multiple regression analysis was used to determine the importance of hemodynamics, surgical injury, and the presence of PTFE on the development of IHT. IHT measurements are expressed at each position as the mean value of the seven dogs with the corresponding standard deviation. The shortest distance along the curved surface of the anastomosis from the suture line to each measurement location (d sl ) was measured. Material 共m兲 was recorded as either zero for artery, or one for PTFE. In vitro wall shear stress was scaled to in vivo values, and the reciprocal was recorded as 1/␶. The multiple regression model was: IHT⫽A 1 共 1/␶ 兲 ⫹A 2 m⫹A 3 共 1/d sl 兲 ⫹B 1 共 m/ ␶ 兲 ⫹B 2 共 1/共 d sl ␶ 兲兲 ⫹B 3 共 m/d sl 兲 ⫹C 1 共 1/␶ 兲 2 ⫹C 2 共 1/d sl 兲 2 ⫹D The coefficients A i represent the linear terms, the B i represent the cross terms, and the C i represent the quadratic terms. There is no C 3 m 2 term because this would be identical to the A 2 m term. The reciprocals of ␶ and d sl are used because IHT is expected to decrease with increasing values of both ␶ and d sl and to approach zero as ␶ and d sl approach infinity. A linear model would yield negative predicted values for IHT, which is not realistic. The linear model could be modified by assuming that all values of IHT are zero beyond the x-intercept, but this would require a piecewise model that is not represented in the simple least squares fit. A drawback to the use of reciprocals is that IHT unrealistically approaches zero at the suture line and for zero. However, in the present case no data was obtained either at the suture line or at zero ␶. Variables are deleted from this model through backward elimination 关26兴. In this procedure, the regression is first performed with all coefficients considered. The coefficient with the Journal of Biomechanical Engineering

Fig. 6 Histomicrographs of graft hood and arterial floor IHT „Ã120 mag…, cross-sections were stained using the Gomoritrichrome-aldehyde fuchsin procedure for connective tissue staining. Graft hood IHT „Fig. 6„a…… consists of an intimal pannus morphologically consistent with smooth muscle cells overlying the PTFE graft. There is no evidence of mural or luminal thrombosis. Floor IHT is represented in Fig. 6„b… „Ã120 mag….

FEBRUARY 2002, Vol. 124 Õ 47

Table 1 Linear regression analyses of various hemodynamic variables versus mean IHT at all locations and the different anastomotic regions

Fig. 7 Distribution of mean IHT at the different axial locations in the experimental end-to-side anastomosis. IHT is along the graft hood and heel is more prominent than other regions within the anastomosis. IHT was absent on the floor of the graft.

within a relatively abundant extracellular matrix. The graft hood and artery floor IHT were similar in structural composition 共Figs. 6共a兲 and 6共b兲, respectively兲. There was no evidence of fresh or organized thrombi. Quantitative Morphometry of IHT. The distribution of mean IHT at the different axial locations is shown graphically in Fig. 7. Quantitative morphometry demonstrated that spatially averaged IHT was substantial along the graft hood and the suture line at the toe and heel 共positions ⫺0.4 and ⫺5.2, respectively兲, but was absent along the floor of all of the digitized sections. IHT was present along the heel side of the POS but absent on the floor side. IHT involving the native artery ranged from 0.0⫾0.1 mm to 0.05 ⫾0.03 mm. A greater amount of IHT was found on the graft hood 共PTFE兲 ranging from 0.09⫾0.06 to 0.24⫾0.06 mm between the toe 共position ⫺0.4兲 and position ⫺4.4. Significant IHT was also found on the graft-side of the heel ranging from 0.06⫾0.04 to 0.17⫾07 mm.

arteries 共10–20 dynes/cm2兲 as a consequence of the relatively large graft diameter and the division of flow in the DOS and POS. Higher values of wall shear stress were located along the floor just proximal to the toe region and within the DOS. PS was significantly higher near the floor stagnation point 共positions ⫺2.0, ⫺2.8, and ⫺3.6兲 than at the same axial positions along the hood 共39⫾0.5 versus 24⫾5, p⬍0.03 respectively兲. However, values of the oscillatory shear index on the floor and hood were not significantly different 共0.17⫾0.02 versus 0.22⫾0.02, p⬎0.1 respectively兲.

II Wall Shear Stress Measurements. Low hemodynamic wall shear stress was found on the hood of the anastomosis and near the flow stagnation region along the floor 共Fig. 8兲. Along the hood, mean wall shear decreased to a minimum value of 1.2 dynes/cm2 at a position two diameters proximal to the toe. On the floor, where flow divided into the proximal and distal directions, the measured minimum was located 2.8 diameters proximal to the toe and with magnitude 1.6 dynes/cm2. At most positions, the wall shear stress was lower than the value typically cited for normal

Fig. 8 Time-averaged wall shear stress for each axial measurement location at the midplane

48 Õ Vol. 124, FEBRUARY 2002

Fig. 9 Linear regression of the reciprocal of mean wall shear „1ÕWSS… and mean IHT. Filled symbols are along the floor „native artery…. Open symbols are along the hood „PTFE…. The triangles are near the suture line. The diamond is the floor stagnation point. Locations other than the suture line and stagnation point are represented with boxes. While a correlation is seen between IHT and 1ÕWSS, the degree of correlation is reduced by the influence of material and injury. The IHT values along the native artery are generally lower than those along PTFE for a given level of shear stress. Also, IHT values near the suture line are higher than those away from the suture line for a given material and a given wall shear stress level.

Transactions of the ASME

Fig. 10 Comparison of predicted IHT by nonlinear multiple regression with backward elimination with measured IHT. The symbols are identical to those used in Fig. 7.

III Quantitative Correlation Between Wall Shear Stress and Anastomotic IHT. Multiple correlations between wall shear stress and anastomotic IHT were examined to assess the role of wall shear stress at the various regions where IHT developed. All 24 anastomotic measurement positions were used to compute linear regressions between IHT and the hemodynamic parameters: mean shear, maximum shear, minimum shear, reciprocal of mean shear, reciprocal of maximum shear, reciprocal of minimum shear, OSI and PS 共Table 1兲. The results revealed a significant correlation only for the reciprocal of mean shear 共r⫽0.525, p⬍0.01, Fig. 9兲 and the reciprocal of maximum shear stress (r⫽0.42, p ⬍0.05). The nonlinear multivariable regression indicated significance for four variables, 1/d sl , (p⫽0.003), m/d sl , (p⫽0.004), m/ ␶ 2 (p⫽3.8⫻10⫺8 ), and 1/d sl , (p⫽0.006). This indicates a model for IHT of the form: IHT⫽0.033/d sl ⫹0.27(m/ ␶ ) ⫹0.0092(m/d sl )⫺0.0019(1/d sl ) 2 ⫺0.060. Units for these variables are IHT 共mm兲, d sl (cm), ␶ 共dynes/cm2兲, and m 共nondimensional兲. The measured IHT is plotted as a function of the predicted IHT from this model in Fig. 10. A linear regression analysis between predicted and measured IHT gives a good correlation coefficient (r⫽0.936, p⫽1.9⫻10⫺11).

Discussion It is reasonable to suspect that hemodynamic stress plays a role in the development of intimal hyperplasia. Both animal and human studies have shown that arteries adjust their lumen diameter in response to long term changes in blood flow rate to maintain the mean wall shear stress between 10 to 20 dynes/cm2 关13–15兴. Atherosclerotic plaques have been shown to localize in low or oscillatory wall shear stress regions of the human carotid artery 关30,32,33兴, the coronary arteries 关34 – 40兴, and the abdominal aorta 关41兴. Shear stress has been shown to elicit specific matrix synthesis responses in smooth muscle cells and to inhibit smooth muscle cell proliferation 关42,43兴. Although these studies indicate that hemodynamics can play a role in adjustment of lumen caliber and in the pathogenesis of disease, they do not specifically address the type of vascular adaptation or growth found in bypass graft anastomoses. It is not known whether significant point to point variation in wall shear stress magnitude within a geometric configuration such as the vascular anastomosis could result in corresponding focal variations in nonatherosclerotic IHT. FurtherJournal of Biomechanical Engineering

more, effects of other confounding variables may be so strong as to mask the effect of wall shear stress. Most investigations into the relationship between hemodynamics and local wall adaptation in end-to-side anastomosis models have not used quantitative intimal thickness data to correlate with wall shear stress 关9,17– 19兴. The studies of Geary et al. 关16兴 were quantitative, but they considered neointimal thickness in a relatively straight segment of PTFE, where spatial variations in hemodynamic wall shear stress are minimal compared to those within a distal end-to-side anastomosis. The results reported herein are unique in that the in vivo canine model and in vitro flow model were developed concurrently. The in vivo model was standardized to maintain similar PTFE diameter and arteriotomy length in all of the animals. The in vitro flow model was designed to match the in vivo graft shape, dimensions, and flow conditions. This design allowed an objective point by point comparison of IHT in the in vivo model with hemodynamic wall shear stress in the in vitro model. The results demonstrate that a significant inverse correlation exists between wall shear stress and anastomotic IHT on the PTFE surface. The degree of correlation suggests that in the presence of PTFE, mean wall shear stress plays a role in the development of anastomotic IHT in addition to other factors, such as mechanical stresses and mural injury. The data suggest that the intimal hyperplastic response to hemodynamic forces may differ in native arteries and prosthetic conduits. IHT in the native artery 共filled symbols in Fig. 9兲 never exceeded 0.05 mm, whereas most measurements along the PTFE 共open symbols in Fig. 9兲 were greater than this value. This may be related to phenotypic heterogeneity 关44兴 in smooth muscle cells at these locations. The differential response of different smooth muscle phenotypes to specific levels of oscillatory wall shear deserves further investigation. Curve fits with hemodynamic variables other than mean wall shear stress 共maximum WSS, minimum WSS, OSI, PS, and the reciprocal of WSS maximum and minimum兲 produced smaller correlation coefficients and larger p-values. Thus, the relatively poor correlation with hemodynamics was not a result of selecting the wrong hemodynamic variable. Wall shear stress gradient was not examined in these correlations because only a rough estimate of the wall shear stress gradient can be obtained from measurements obtained in this study. A numerical study of this flow model, validated by the in vitro data, should provide sufficient shear stress gradient data to perform quantitative correlations. There are several differences between the hemodynamics of the hood and the floor. The pulse shear was significantly higher along the floor than along the hood, but the oscillating shear index was not. Additionally, reversal of shear stress along the hood represents true flow separation, similar to that which occurs at the carotid artery bulb. In contrast, changes in direction of flow along the floor are caused by a division of flow between the proximal and distal outflow segments 共stagnation flow region兲 and occur without separation. Because the IHT was always contiguous with the suture line, and because there was a general decrease in IHT with d sl , this study suggests that IHT is initiated along the suture line and migrates onto both artery and PTFE. It appears that this migration of IHT was not rapid enough to reach the floor within the three month period of this study 共i.e. more rapid migration of IHT on the PTFE surface兲. An important question is whether a longer study would allow this migration to reach the floor and whether a stronger correlation with 1/␶ would then be found along the native artery. The parameters in the final curve fit 共Table 2兲 are notable. After elimination of terms that were poorly correlated, the only term remaining with mean wall shear stress was the cross-term with material (m/ ␶ ). Since host artery is represented as m⫽0, only shear stress in conjunction with PTFE produced a statistically significant correlation 共i.e. no statistically significant correlation was FEBRUARY 2002, Vol. 124 Õ 49

Table 2 Nonlinear multivariable backward elimination results

produced with native artery IHT兲. The term 1/d sl is represented by both a linear and a nonlinear term. Material is represented only in combination with d sl and ␶, which suggests a strong coupling between material and the other two parameters. It must be recognized that the dependencies on ␶ and on d sl have singularities at ␶⫽0 and at the suture line, respectively. The data used in the curve fit are sufficiently far away from these regions to avoid the singularities. For example, the dependence on d sl peaks at about d sl ⫽0.09 cm and the closest positions to the suture line are at d sl ⫽0.12 cm. The model should not be extrapolated to values of d sl or ␶ smaller than those used in the curve fit. Since IHT was small in all of the canines studied, there may be differences between the intimal thickening observed in the present study and that observed in human vascular anastomoses. Furthermore, the multivariable curve fit model is mathematical rather than physiological. Clearly a model must be developed in which the coefficients are directly related to physiological parameters such as rate of cellular migration, apoptosis and proliferation. However, the data presented here provide an initial basis for comparison with other experiments, and they could be used as an initial test data set for more physiological models. With advances in computers, numerical simulations for each individual animal graft geometry and flow condition may be possible for future studies. This may be important since it is difficult to predict how small changes in geometry may impact the wall shear stress distribution in these complex three-dimensional geometries. Further, a single flow waveform and geometry may not have a wall shear stress distribution representative of each individual graft, which may impact the correlation results.

Conclusion Quantitative and concurrent experiments have been carried out to determine the correlation between intimal hyperplastic thickening in vivo and the distribution of hemodynamic wall shear stress in vitro for a canine end-to-side anastomosis model. The spatial distribution of IHT and the reciprocal of mean wall shear stress were correlated (r⫽0.525, p⬍0.01). This correlation was only present for positions located on the PTFE graft itself. The data indicate that IHT is most evident at the suture line and along the juxta anastamotic PTFE graft hood, in contrast to the native artery. The relationship between IHT development and hemodynamics is significantly affected by multiple biological and mechanical factors 共i.e., hemodynamics, surgical injury, and the presence of PTFE兲. A mathematical model based on multivariable analysis can be used to describe this relationship.

Acknowledgments This work was supported by the National Institutes of Health Grant 5 R01-H2-41267 and the National Heart, Lung and Blood Institute of the United States Public Health Service R01HL55296-01.

References 关1兴 Imparato, A. M., Bracco, A., Kim, G. E., and Zeff, R., 1972, ‘‘Intimal and Neointimal Fibrous Proliferation Causing Failure of Arterial Reconstruction,’’ Surgery, 72, pp. 1007–1017. 关2兴 Echave, V., Koornick, A., Haimov, M., and Jacobson, J., 1979, ‘‘Intimal Hyperplasia as a Complication of the Use of the Polytetrafluoroethylene Graft for Femoral-Popliteal Bypass,’’ Surgery, 86, pp. 791–798.

50 Õ Vol. 124, FEBRUARY 2002

关3兴 Pedrini, L., Pisano, E., Donato, Di Paola M., Balleste, A., and Magnoni, F., 1994, ‘‘Late Occlusion of Aortofemoral Bypass Graft: Surgical Treatment,’’ Cardiovasc. Surg., 2, pp. 763–766. 关4兴 Zempo, N., Esato, K., O-Hara, M., Fujioka, K., Kuga, T., and Takenaka, H., 1993, ‘‘Is the Preferential Use of Polytetrafluoroethylene Grafts for BelowKnee Femoropopliteal Bypass Justified?’’ Int. Surg., 78, pp. 162–165. 关5兴 Bryan, A. J., and Angelini, G. D., 1994, ‘‘The Biology of Saphenous Vein Graft Occlusion: Etiology and Strategies for Prevention,’’ Curr. Opin. Cardiol., 9, pp. 641– 649. 关6兴 Waller, B. F., Pinkerton, C. A., and Foster, L. N., 1987, ‘‘Morphologic Evidence of Accelerated Left Main Coronary Artery Stenosis: A Later Complication of Percutaneous Transluminal Balloon Angioplasty of the Proximal Left Anterior Descending Coronary Artery,’’ J. Am. Coll. Cardiol., 9, pp. 1019– 1023. 关7兴 Liu, M. W., Roubin, G. S., and King, S. B., 1989, ‘‘Restenosis After Coronary Angioplasty: Potential Biologic Determinants and Role of Intimal Hyperplasia,’’ Circulation, 79, pp. 1374 –1387. 关8兴 Glagov, S., 1994, ‘‘Intimal Hyperplasia, Vascular Modeling and the Restenosis Problem,’’ Circulation, 89, pp. 2888 –2891. 关9兴 LoGerfo, F. W., Soncrant, T., Teel, T., and Dewey, Jr., C. F., 1979, ‘‘Boundary Layer Separation in Models of Side-to-End Arterial Anastomoses,’’ Archives of Surgery , 114, pp. 1369–1373. 关10兴 Clark, R. E., Apostolou, S., and Kardos, J. L., 1976, ‘‘Mismatch of Mechanical Properties as a Cause of Arterial Prosthesis Thrombosis,’’ Surg. Forum, 27, pp. 208 –210. 关11兴 LoGerfo, F. W., Quist, W. C., Nowak, M. D., Crawshaw, H. M., and Haudenschild, C. C., 1983, ‘‘Downstream Anastomotic Hyperplasia. A Mechanism of Failure in Dacron Arterial Grafts,’’ Ann. Surgery, 197, pp. 479– 483. 关12兴 Clowes, A. W., Gown, A. M., Hanson, S. R., and Reidy, M. A., 1985, ‘‘Mechanisms of Arterial Graft Failure, Role of Cellular Proliferation in Early Healing of PTFE Prostheses,’’ Am. J. Pathol., 118, pp. 43–54. 关13兴 Kamiya, A., and Togawa, T., 1980, ‘‘Adaptive Regulation of Wall Shear Stress to Flow Change in the Canine Carotid Artery,’’ Am. J. Pathol., 239, pp. H14 – H21. 关14兴 Langille, B. L., and O’Donnell, F., 1986, ‘‘Reductions in Arterial Diameter Produced by Chronic Decreases in Blood Flow are Endothelium-Dependent,’’ Science, 231, pp. 405– 407. 关15兴 Zarins, C. K., Zatina, M. A., Giddens, D. P., Ku, D. N., and Glagov, S., 1987, ‘‘Shear Stress Regulation of Artery Lumen Diameter in Experimental Atherogenesis,’’ J. Vasc. Surg., 5, pp. 413– 420. 关16兴 Geary, R. L., Kohler, T. R., Vergel, S., Kirkman, T. R., and Clowes, A. W., 1994, ‘‘Time Course of Flow-Induced Smooth Muscle Cell Proliferation and Intimal Thickening in Endothelialized Baboon Vascular Grafts,’’ Circ. Res., 74, pp. 14 –23. 关17兴 Sottiurai, V. S., Yao, J. S. T., Batson, R. C., Sue, S. L., Jones, R., and Nakamura, Y. A., 1989, ‘‘Distal Anastomotic Intimal Hyperplasia: Histopathologic Character and Biogenesis,’’ Ann. Vasc. Surg., 24, pp. 711–722. 关18兴 Ojha, M., Ethier, C. R., Johnston, K. W., and Cobbold, R. S., 1990, ‘‘Steady and Pulsatile Flow Fields in an End-to-Side Arterial Anastomosis Model,’’ J. Vasc. Surg., 12, pp. 747–753. 关19兴 Ojha, M., 1994, ‘‘Wall Shear Stress Temporal Gradient and Anastomotic Intimal Hyperplasia,’’ Circ. Res., 74, pp. 1227–1231. 关20兴 White, S. S., Zarins, C. K., Giddens, D. P., Bassiouny, H. S., Loth, F., Jones, S. A., and Glagov, S., 1993, ‘‘Hemodynamic Patterns in a Model of End-to-Side Vascular Graft Anastomoses: Effect of Pulsatility, Flow Division and Reynolds Number and Hood Length,’’ ASME J. Biomech. Eng., 115, pp. 105–111. 关21兴 Keynton, R., Chapman, M. M., Simms, R. L., Rodway, N., and Rittgers, S. E., 1995, ‘‘Direct Relationship between Wall Shear Rate and Intimal Hyperplasia in Vascular Bypass Grafts,’’ ASME BED Advances in Bioengineering, Vol. 31, pp. 169–170. 关22兴 Bassiouny, H. S., White, S., Glagov, S., Choi, E., Giddens, D. P., and Zarins, C. K., 1991, ‘‘Anastomotic Intimal Hyperplasia: Mechanical Injury or Flow Induced,’’ J. Vasc. Surg., 15, pp. 708 –717. 关23兴 Bassiouny, H. S., Krievins, D., Glagov, S., Abu-hamid, G., and Zarins, C. K., 1993, ‘‘Distal Arteriovenous Fistula Inhibits Experimental Anastomotic Intimal Thickening,’’ Surg. Forum, XLIV, pp. 345–346. 关24兴 Giddens, E. M., Giddens, D. P., White, S. S., Zarins, C. K., Bassiouny, H. S., and Glagov, S., 1993, ‘‘Exercise Flow Conditions Eliminate Stasis at Vascular Graft Anastomoses,’’ Proceedings of the Third Mid-Atlantic Conference in Biofluid Mechanics, pp. 255–267. 关25兴 Bassiouny, H. S., Ng, A., and Glagov, S., ‘‘Low Shear Stress Enhances Progression of Anastomotic Intimal Hyperplasia,’’ Submitted to Ann. Vasc. Surg. 关26兴 Cleary, J. P., and Levenbach, H., 1982, The Professional Forecaster: The Forecasting Process Through Data Analysis, Lifetime Learning Publications, Belmont, Ca. 关27兴 Loth, F., 1993, ‘‘Velocity and Wall Shear Measurements Inside a Vascular Graft Model Under Steady and Pulsatile Flow Conditions,’’ Ph.D. Thesis, Georgia Institute of Technology, Atlanta, GA. 关28兴 Loth, F., Jones, S. A., Giddens, D. P., Bassiouny, H. S., Zarins, C. K., and Glagov, S., 1997, ‘‘Measurements of Velocity and Wall Shear Stress Inside a PTFE Vascular Graft Model Under Steady Flow Conditions,’’ ASME J. Biomech. Eng., 119, pp. 187–194. 关29兴 Loth, F., Jones, S. A., Giddens, D. P., and Brossollet, L. J., 1994, ‘‘Accuracy of Wall Shear Stress Estimates from Laser Doppler Anemometry Measurements Under Unsteady Flow Conditions,’’ Advances in Bioengineering, Proceedings of the ASME Winter Annual Meeting, pp. 307–308. 关30兴 Ku, D. N., Giddens, D. P., Zarins, C. K., and Glagov, S., 1985, ‘‘Pulsatile Flow

Transactions of the ASME

关31兴

关32兴

关33兴 关34兴 关35兴 关36兴 关37兴 关38兴

and Atherosclerosis in the Human Carotid Bifurcation: Positive Correlation Between Plaque Location and Low and Oscillating Shear Stress,’’ Arteriosclerosis 5, pp. 293–302. Glagov, S., and Zarins, C. K., 1983, ‘‘Quantitating Atherosclerosis; Problems of Definition in Clinical Diagnosis,’’ Clinical Diagnosis of Atherosclerosis Quantitative Methods of Evaluation, M. Bond et al., eds., Springer-Verlag, New York, pp. 11–35. Zarins, C. K., Giddens, D. P., Bharadvaj, B. K., Sottiurai, V. S., Mabon, R. F., and Glagov, S., 1983, ‘‘Carotid Bifurcation Atherosclerosis: Quantitative Correlation of Plaque Localization With Flow Velocity Profiles and Wall Shear Stress,’’ Circ. Res., 53, pp. 502–514. Bharadvaj, B. K., Mabon, R. F., and Giddens, D. P., 1982, ‘‘Steady Flow in a Model of the Human Carotid Bifurcation: I. Flow Visualization,’’ ASME J. Biomech. Eng., 15, pp. 348 –362. Svindland, A., 1983, ‘‘The Localization of Subanophilie and Fibrous Plaques in the Main Left Coronary Arteries,’’ Atherosclerosis 48, pp. 139–145. Fox, B., James, K., Morgan, B., and Seed, A., 1982, ‘‘Distribution of Fatty and Fibrous Plaques in Young Human Coronary Arteries,’’ Atherosclerosis 41, pp. 337–347. Sabbah, H. N., Khaja, F., Brymer, J. F., Hawkins, E. T., and Stein, P. D., 1984, ‘‘Blood Velocity in the Right Coronary Artery: Relation to the Distribution of Atherosclerotic Lesions,’’ Am. J. Cardiol., 53, pp. 1008 –1012. Friedman, M. H., Bargeron, C. B., Deters, O. J., Hutchins, G. M., and Mark, F. F., 1987, ‘‘Correlation Between Wall Shear and Intimal Thickness at a Coronary Artery Branch,’’ Atherosclerosis 68, pp. 27–33. Tang, T., 1990, ‘‘Periodic Flow in a Bifurcating Tube at Moderate Reynolds

Journal of Biomechanical Engineering

Number,’’ Ph.D. Thesis, Georgia Institute of Technology, Atlanta, GA. 关39兴 Asakura, T., and Karino, T., 1990, ‘‘Flow Patterns and Spatial Distribution of Atherosclerotic Lesions in Human Coronary Arteries,’’ Circ. Res., 66, pp. 1045– 66. 关40兴 Krams, R., Wentzel, J. J., Oomen, J. A. F., Vinke, R., Schuurbiers, J. C. H., Feyter, P. J., Serruys, P. W., and Stager, C. J., 1997, ‘‘Evaluation of Endothelial Shear Stress and 3D Geometry as Factors Determining the Development of Atherosclerosis and Remodeling in Human Coronary Arteries in Vitro, Combining 3D Reconstruction from Angiography and IVUS 共ANGUS兲 With Computational Fluid Dynamics,’’ Arterioscler., Thromb., Vasc. Biol., 17, pp. 2061– 2065. 关41兴 Moore, Jr., J. E., Xu, C., Glagov, S., Zarins, C. K., and Ku, D. N., 1994, ‘‘Fluid Wall Shear Stress Measurements in a Model of the Human Abdominal Aorta; Oscillatory behavior and relationship to Atherosclerosis,’’ Atherosclerosis 110, pp. 225–240. 关42兴 Papadaki, M., McIntire, L. V., and Eskin, S. G., 1996, ‘‘Effects of Shear Stress on the Growth Kinetics of Human Aortic Smooth Muscle Cells in Vitro,’’ Biotechnol. Bioeng., 50, pp. 555–561. 关43兴 Alshihabi, S. N., Chang, Y. S., Frangos, J. A., and Tarbell, J. M., 1996, ‘‘Shear Stress-Induced Release of PGE2 and PGI2 by Vascular Smooth Muscle Cells,’’ Biochem. Biophys. Res. Commun., 224, pp. 808 – 814. 关44兴 Gadson, Jr., P. F., Rossignol, C., McCoy, J., and Rosenquist, T. H., 1993, ‘‘Expression of Elastin, Smooth Muscle Alpha-Actin, and C-Jun as a Function of the Embryonic Lineage of Vascular Smooth Muscle Cells,’’ In Vitro Cell. Dev. Biol.: Anim., 29A, pp. 773–781.

FEBRUARY 2002, Vol. 124 Õ 51