ULTRASONIC FATIGUE TEST OF AZ91 MAGNESIUM ALLOY

4 downloads 0 Views 1MB Size Report
sensor is transformed to pulses and connected to counter input of NI. 9401 module. Temperature is monitored with a thermocouple connected to NI 9211 module ...
MACHINES, TECHNOLOGIES, MATERIALS. ISSN 1313-0226. ISSUE 11/2013

ULTRASONIC FATIGUE TEST OF AZ91 MAGNESIUM ALLOY D. Dimitrov PhD. 1, V. Shtarbakov PhD. 2 Technical University-Varna, Bulgaria, 9000 Varna, 1 Studentska str.1 Metal PLC-Varna, Western Industrial Zone 2 Abstract: In this article questions about fatigue endurance limit of Az91 Mg-alloy was discussed. Very high cycle fatigue testing at 20kHz (push-pull mode) was done. An ultrasonic testing system developed in Technical University of Varna was used. Questions about fracture micromechanisms and crack initiations were also discussed. Keywords: VHCF, MG-ALLOY, FATIGUE ENDURANCE LIMIT;

generator can deliver 1kW power in a wide frequency range (15100kHz). The transducer with power up to 3kW, has working (resonance) frequency 20±0.5kHz and can produce maximum peakto-peak amplitude 20μm. The ultrasonic generator has closed loop control of supplied voltage and frequency in order to keep the resonance conditions and the constant vibration amplitude. The vibrations of the transducer are amplified by a booster (3) and a sonotrode (4). The booster have amplification ratio 1:1.5, resonance frequency around 20kHz and is made of Ti6Al4V-alloy. It’s designed with a wide flange in nodal plane for fixing the whole system. The sonotrode has a conical shape with amplification ratio around 1:3 and is made of Al7075-T6 alloy [18]. Symmetrical specimens with stress concentrator in the middle (hourglass or dumbbell) have to be designed. The length of the specimen is calculated from the half wave (λ/2) condition [19] i.e. center of the specimen is displacement node and ends are displacement antinodes, fig.1. Measurement system The measurement system consists of a displacement sensor and a thermocouple. NI cDAQ chassis with 9211, 9215, and 9401 modules is used for data acquisition. The displacements at the end of the specimen are monitored with eddy-current displacement sensor connected to NI9215 module. For cycle counting, the signal from eddy current sensor is transformed to pulses and connected to counter input of NI 9401 module. Temperature is monitored with a thermocouple connected to NI 9211 module. The hot end of the thermocouple is glued to the center of the specimen. Cooling system A circular cooling system was designed for preventing specimen heating during operation. Drilling oil as a cooling agent is used.

1.Introduction Mg-alloys are known as materials with high specific strength and damping capacity. They have also good castability and are easy to recycle. Recently, they attract increased attention as possible structural material in aerospace and automotive industries [1]. Of course, their limited ductility, easy oxidation and relatively low ultimate strength, constrain for now, their wide applications. Researches of Mg-alloys concentrate their investigations mainly on the following topics: corrosion resistance [2, 3, 4, 5]; creep resistance [6, 7, 8], and fatigue life durability [9, 10, 11, 12, 13,14]. Among Mg-alloys AZ91 is one of the most used (more than 50% of all Mg-alloys). The aim of the current investigation is to obtain fatigue endurance strength of this alloy on a base of 108 cycles and to analyze broken surfaces. For light metals, it is known that there is no true fatigue limit. However, the fatigue life of current automobile engines ranges around 108 cycles, around 109 or even 1010 cycles are in interest for turbine engine components [10,15]. Such experiments for conventional fatigue machines are very time consuming, but if ultrasonic fatigue testing at frequency 20kHz is used, 108 cycles can be reached in about two hours.

2.Ultrasonic fatigue testing and equipment To develop high stresses in the tested specimen ultrasonic fatigue testing requires cycle loading at its resonance frequency. For this purpose all vibrating parts of the testing machine, as well as the specimen, have to be designed with appropriate shape and dimensions. Ultrasonic fatigue machines usually have three main systems: an ultrasonic system, which consists of a generator, a piezoceramic transducer, an amplification unit, and a specimen and supply resonance mode vibrations at 20kHz; a measuring and control system for measuring vibration amplitudes, cycles counting and a generator “on-off” control; additional systems like a cooling device, a crack growth measurement device, etc. Ultrasonic system Scheme of ultrasonic system of testing machine, designed in TU-Varna is shown on fig.1 [16] The generator (1) delivers high voltage sinusoidal signal to the piezoceramic transducer (2), which converts the electrical signal to mechanical vibrations. Both devices are commercially available from MPI-Switzerland, designed for ultrasonic welding, but fit well the requirements of fatigue testing [17]. The

Fig.1 Ultrasonic fatigue machine designed and assembled in Strength of Materials lab of Technical University of Varna. 1.Generator; 2Ttransducer; 3Booster ( MPInerconsulting – Switzerland) 4-Conical shape sonotrode; 5.Test sample.

50

MACHINES, TECHNOLOGIES, MATERIALS. ISSN 1313-0226. ISSUE 11/2013

from ingot and machined. Impulse-resonance method according to ASTM E1876 [20], was applied. Mean value for dynamic Young’s Specimen Preparation modulus, obtained by exiting fundamental longitudinal mode of the Cast Mg-alloy AZ91 (Mg-9Al-1Zn) with chemical samples, is E=42.45GPa. composition given in table.1 is used for this study. Specimens were Fatigue Calibration Chemical composition of investigated alloy AZ91, wt % Table.1 Calibration was done with cylindrical test specimen Chem. Al Cu Fe Mn Ni Si Zn with λ/2 resonance length (L=121mm). Since applied voltage is Element linearly proportional to displacement amplitude at the end and * stress in the middle of the specimen, they are simultaneously Min., 8.1 0.17 0.4 measured using test equipment shown on fig.3. Stress is % measured with strain gauge (1) mounted in the middle of the specimen and displacement amplitude by gauge indicator (2) Max., 9.3 0.015 0.005 0.35 0.001 0.2 1 % with 0.001mm precision. *-Mg- bal. Fatigue testing For fatigue testing 108 cycles are chosen as an endurance limit. Specimens were constantly cooled in order to keep temperature around 25oC. Stress amplitude is controlled by changing voltage to the piezoceramic transducer. Ø5

3. Experimental Procedure

Ø10

4. Results and Discussion

Ultimate tensile strength Rm=167MPa and elongation A=2,9%, were determined from static tensile test. These values are in the range of standard for such alloys [21]. Microstructure consists of α-Mg, eutectic α+γ and discontinuous precipitations of γ-phase (Mg17Al12), fig.4. Fatigue life result, in the range 105-108 cycles are presented in the form of S-N diagram, fig.4. Fatigue limit at 108 cycles is in the range 44-55MPa. Using staircase method, mean fatigue strength is estimated to σf50=49.8MPa, standard deviation to Sd=5.5MPa, so lower fatigue strength estimated on 90% reliability is σf90=40.4MPa. Such values for fatigue strength are similar to those obtained from researches in [9, 10, 12, 14]. In the fatigue failure zone (fig.5) dispersion of the results is high, so adequate fatigue equation cannot be obtained. One of the possible reasons for this is the presence of cast defects in tested samples (voids, pores, structural inhomogeneity). Presence of such defects, their nature, size, orientation and distribution, are detrimental to fatigue endurance of specimens. Experimental values can be divided to two regions. The most likely reasons for fatigue crack initiation and development in terms of specimens failed in the first region (I- 105-106 cycles), is a presence of internal voids. However region “I” may be divided into two subregions: a broken at relatively high stress specimens (A), and broken at lower load values specimens (B). In specimens of group (A) sites of the fatigue cracks initiation are predominantly close to the surface voids. The possibility of this is determined by the high values of the applied stress amplitude in the order of 70-90 MPa. This is proved from the broken surface shown on fig.6. Crack origins are shown with arrows. Relatively low cycle crack development in this case leads to formation of rough α+γ fracture surface. Static fracture α - Mg zone is missing. For specimens γMg17Al12 of group (B) main precipitations 8 µm reason for fatigue fracture is Fig.4 Microstructure of investigated cast Mg-Alloy AZ91 presence

Fig.2 Test specimen from cast Mg-Alloy AZ91.

cut from ingot, machined and polished. For the fatigue test “hourglass” specimens with appropriate shape and dimensions were prepared, fig.2 Diameters of cylindrical part and the throat, as well as length of reduced part were technologically chosen. Analytical solution given in [15] of the longitudinal wave equation for obtaining the resonance length and radius of reduced part is used. Dimensions from analytical solution were verified by FEM frequency modal analysis [19]. It has to be mentioned that for proper design of the specimen exact values for Dynamic Young’s Modulus and density have to be obtained. Since experimentally measured density 1 (Archimedes’ method) is ρ=1812kg/m3, practically 2 identical with values given in the booklets, critical parameter is Young’s modulus which has to be determined experimentally for the investigated pieces of 3 material. In this study samples Fig.3. Calibration of stresses displacement cylindrical amplitude 1-test specimen AZ9; 2-strain (ф10x100) for dynamic gauge; 3-gauge indicator Young’s modulus measurement was cut

51

MACHINES, TECHNOLOGIES, MATERIALS. ISSN 1313-0226. ISSUE 11/2013

As for the experimental results of the second area (“II”-107-108, fig.5.), the most likely reason of fatigue fracture is the presence of structural heterogeneity. Microstructure of the alloy from which the samples are made, shown on fig. 3 is conformation. The presence of clearly visible discontinuous γ-phase precipitations is possible reason for the initiation and development of fatigue crack in high-cycle fatigue region. Fractured surface of specimen cycled with stress amplitude σa=55MPa and failed at N=5.42x107 cycles is shown on fig.7. The fatigue crack propagation area (zone II) and the static fracture (zone I) are clearly visible. The crack origin is at the surface. Probably fatigue cracks developed from microcracks initiated at the phase border between γphase (Mg17Al12) precipitations and the main α-Mg matrix. Heat treatment to dissolve these precipitations can improve strength of the alloy [10].

A

B

I

II

Fig.5. S-N plot of cast Mg-alloy AZ91 σf50=49.8MPa σf90=40.4MPa

5.Conclusions Based on the ultrasonic fatigue test at frequency 20KHz in “push-pull” mode with symmetrical cycle loading (R=-1), of the fatigue life of the studied magnesium alloy AZ91 in the range 105108 cycles, following main conclusion can be mentioned: - the mechanism of fatigue damage and the nature of fatigue fracture are strictly dependent on the structural condition of the alloy (microstructure, presence of cast defects); - if there is casting defects as voids or internal pores, they act as a crack initiation sites and fatigue life is up to 106 cycles. Depending on size and distribution of voids specimens can fail at high or low amplitude stresses; -if there is no large voids fatigue life exceeds 107 cycles at amplitude stress less than 80MPa. Fatigue fracture origins are on the surface, probably initiated at the phase border between γ-phase (Mg17Al12) precipitations and the main α-Mg matrix. The fatigue crack propagation area and the static fracture are clearly visible on fracture surface; -on a base of 108 cycles mean fatigue strength of the investigated alloy is σf50=49.8MPa

0.5 mm

Acknowledgements Ultrasonic fatigue machine used for the experiments was developed with the financial support of Bulgarian “Science Research” fund under the project DMU03/98.

Fig.6. Fractured surface after 3*105 cycles, 65 MPa loading.

References [1] R.W.Cahn, C.X.Shi and J.Ke – Structure and Properties of Nonferrous Alloys, Beijing, Science Press, 1999, p.5564 [2] Yong-Gang Li, Ying-Hui Wei, Li-Feng Hou, Peng-Ju Han - Atmospheric corrosion of AM60 Mg alloys in an industrial city environment , Corrosion Science, Volume 69, April 2013, p.67-76 [3] A.S. Gnedenkov, S.L. Sinebryukhov, D.V. Mashtalyar, S.V. Gnedenkov - Features of the corrosion processes development at the magnesium alloys surface , Surface and Coatings Technology, Volume 225, 25 June 2013, p.112-118 [4] Yingwei Song, En-Hou Han, Dayong Shan, Chang Dong Yim, Bong Sun You - The effect of Zn concentration on the corrosion behavior of Mg–xZn alloys , Corrosion Science, Volume 65, December 2012, p. 322-330 [5] Sebastián Feliu Jr., C. Maffiotte, Juan Carlos Galván, Violeta Barranco - Atmospheric corrosion of magnesium alloys AZ31 and AZ61 under continuous condensation conditions, Corrosion Science, Volume 53, Issue 5, May 2011, p. 1865-1872 [6] A. Srinivasan, K.K. Ajithkumar, J. Swaminathan, U.T.S. Pillai, B.C. Pai - Creep Behavior of AZ91 Magnesium Alloy, Procedia Engineering, Volume 55, 2013, p. 109-113 [7] Luhong Wang, Haozhe Liu - The microstructural evolution of Al12Mg17 alloy during the quenching processes,

I

II

0.5 mm Fig.7. Fractured surface after 5.42*107 cycles, 55MPa loading.

of large voids or group of voids, which cause high stress concentration leading to fast fracture at relatively low amplitude stresses.

52

MACHINES, TECHNOLOGIES, MATERIALS. ISSN 1313-0226. ISSUE 11/2013

Journal of Non-Crystalline Solids, Volume 352, Issues 26–27, 1 August 2006, p. 2880-2884 [8] Meng Xian Ying, Derek O Northwood, Ian O Smith Microstructures developed during steady-state creep of Al-Mg alloys, Metallography, Volume 19, Issue 3, August 1986, p. 285304 [9] B. Wolf , C. Fleck , D. Eifler - Characterization of the fatigue behaviour of the magnesium alloy AZ91D by means of mechanical hysteresis and temperature measurements, International Journal of Fatigue, 26, (2004), p.1357–1363 [10] Aneta Němcová, Josef Zapletal, Martin Juliš, Tomáš Podrábský - Cyclic fatigue resistance of AZ91 magnesium alloy, Materials Engineering, Vol. 16, 2009, No. 4, p.5-10 [11] Thomas A. Koch1, Michael A. Gharghouri - Fatigue properties of die-cast magnesium alloy AM60B, Magnesium Technology 2003 Edited by Howard I. Kaplan TMS (The Minerals, Metals & Materials Society), 2003, p.71-76 [12] M.F. Horstemeyer , N. Yang, Ken Gall , D.L. McDowell, J. Fan, P.M. Gullett - High cycle fatigue of a die cast AZ91E-T4 magnesium alloy, Acta Materialia 52 (2004), p.1327– 1336 [13] A. N. Chamos, Sp. G. Pantelakis, G. N. Haidemenopoulos, E. Kamoutsi - Tensile and fatigue behaviour of wrought magnesium alloys AZ31 and AZ61, Fatigue & Fracture of Engineering Materials & Structures, 31, 2008, p. 812-821

[14] Yeong-Jern Chen, Bai-Sen Wei , Wen-Nong Hsu Ultrasonic Fatigue Testing of Aluminum and Magnesium Alloys, Journal of the Chinese Society of Mechanical Engineers, Vol.29, No.4, (2008), pp.299-308 [15] Claude Bathias, Paul Croce Paris, Gigacycle Fatigue in Mechanical Practice, Marcel Dekker, 2005 - 304 pages [16] Diyan Dimitrov, Veselin Mihailov, Borislav Kostov, Modeling of Ultrasonic Fatigue-Life Testing Machine, Proceedings of COMSOL conference, Milan 2012, Italy 2012 [17]http://www.mpi-ultrasonics.com/ [18]Диян М. Димитров, Веселин Михайлов, Борислав Костов, Проектиране на сонотрод на лтразвуков стенд за изпитване на умора, Сборник доклади трети международен научен конгрес "50 ГОДИНИ ТУ – ВАРНА", том 4,20120026-0031 [19] Димитров Д. М., Михайлов В., Костов Б., Димитров Р. МОДЕЛИРАНЕ НА ПРОБНИ ОБРАЗЦИ ЗА ИЗПИТВАНЕ НА УМОРА ПРИ ЧЕСТОТА 20kHz, Научни известия на НТСМ , година ХХ, Брой 1(133), Юни 2012, стр 155-159 ,ISSN 1310-3946 [20] ASTM International, Designation: E1876-09, Standart test method for dinamic Young`s modules, Shear modules, and Poasson`s ratio by Impulse Excitation of vibration. [21] http://www.magnesium.com